FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N
- Recall number
- Z-2607-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N
- Status
- Open, Classified
- Initiated
- 2026-05-28
- Posted
- 2026-06-26
- Terminated
- not on file
- Reason
- Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
- Root cause
- Process change control
- Action
- On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 9,272 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
- Lot, serial or model codes
- DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OEZ
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28