Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho Development Corporation: Encompass Hip Stem, Plasma Spray, 12/14 Taper, Standard, REF 108-XXXX, Sterile R, Ortho Development, Draper, Utah 84020. Intended for use in total hip arthroscopy. Device fixation is achieved by an optimal, press-fit in the medullary canal, which maximizes contact between the stem and the bone.

Recall number
Z-2602-2011
Recalling firm
Ortho Development Corporation
Product
Encompass Hip Stem, Plasma Spray, 12/14 Taper, Standard, REF 108-XXXX, Sterile R, Ortho Development, Draper, Utah 84020. Intended for use in total hip arthroscopy. Device fixation is achieved by an optimal, press-fit in the medullary canal, which maximizes contact between the stem and the bone.
Status
Terminated
Initiated
2011-05-10
Posted
2011-06-20
Terminated
2011-11-09
Reason
Tests on PVC packaging material resulted in alert levels for cytotoxicity.
Root cause
Package design/selection
Action
Ortho Development Corporation notified consignees by letter on May 11, 2011. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to return all remaining affected units to Ortho. Replacement product was to be shipped by May 31.
Quantity
113 units
Distribution
Nationwide Distribution - USA including CA, FL, and TX.
Lot, serial or model codes
REF Number 108- 1017, 1019, 1217, 1219, 1221, 1223, 1419, 1421, 1423, 1425, 1623, 1625, 1627, 1825, 1827. All Lot numbers.
510(k) numbers
["K053293"]
PMA numbers
not on file
Product code
MBL
Firm FEI
3017298928
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register