Siemens Healthcare Diagnostics Inc: Siemens RAPIDPoint 500 v2.2.2A Software Upgrade Kit; SMN 11066719, Software Version 2.2A The RAPIDPoint 500 system is designed for professional use in a point-of care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili

Recall number
Z-2601-2016
Recalling firm
Siemens Healthcare Diagnostics Inc
Product
Siemens RAPIDPoint 500 v2.2.2A Software Upgrade Kit; SMN 11066719, Software Version 2.2A The RAPIDPoint 500 system is designed for professional use in a point-of care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili
Status
Terminated
Initiated
2016-07-05
Posted
2016-08-19
Terminated
2017-08-02
Reason
Some v2.2.2 upgrade kits include a dialysate mode which not cleared/approved for shipment in the United States
Root cause
Under Investigation by firm
Action
Siemens sent a Customer Notification dated July 2016 to all affected domestic (US) customers via Federal Express on July 5, 2016. Customers were asked to take the following actions: Please do not use the dialysate mode. If you have enabled this feature, please disable it by following these instructions: 1. At the System screen, select Setup. 2. Select Secured Options > Analysis Options. 3. Unselect Dialysate. 4. Select the Continue button 3 times. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check attached to this letter within 7 days. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens Technical Support representative. For questions regarding this recall call 781-269-3000.
Quantity
76
Distribution
Nationwide Distribution to AL, CA, MA, NJ, OH, TX, WA, PA
Lot, serial or model codes
Software Version 2.2A
510(k) numbers
["K113216"]
PMA numbers
not on file
Product code
KHP
Firm FEI
3002637618
Firm city
Norwood
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register