Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: power cord for: (a) AMIA Automated PD Cycler, Product Code: 5C9320; (b) Kaguya PD System, Product Code: T5C8500 (Japanese distribution only)

Recall number
Z-2597-2018
Recalling firm
Baxter Healthcare Corporation
Product
power cord for: (a) AMIA Automated PD Cycler, Product Code: 5C9320; (b) Kaguya PD System, Product Code: T5C8500 (Japanese distribution only)
Status
Terminated
Initiated
2018-07-12
Posted
not on file
Terminated
2021-03-31
Reason
There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
Root cause
Under Investigation by firm
Action
Method of Notification: A Safety Alert communication will be sent to affected customers via U.S.P.S., first class mail. Mitigating Instructions: Baxter is asking customers to: 1. Unplug the devices from the wall before unplugging the power cord from the back of the cycler. 2. Confirm the receipt of the letter by completing the enclosed customer reply form and returning it to Baxter by either mail, fax, or scanned email.
Quantity
5445 units
Distribution
US and Japan
Lot, serial or model codes
All Serial Numbers
510(k) numbers
["K151525"]
PMA numbers
not on file
Product code
FKX
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register