Biomet 3i, LLC: BIOMET 3i Certain Bellatek Encode Healing Abutment Biomet 3i Healing abutments are temporary abutments indicated for use in maintaining the soft tissue opening throughout the healing process prior to dental restoration.
- Recall number
- Z-2594-2014
- Recalling firm
- Biomet 3i, LLC
- Product
- BIOMET 3i Certain Bellatek Encode Healing Abutment Biomet 3i Healing abutments are temporary abutments indicated for use in maintaining the soft tissue opening throughout the healing process prior to dental restoration.
- Status
- Terminated
- Initiated
- 2014-06-25
- Posted
- 2014-09-05
- Terminated
- 2014-09-23
- Reason
- Identifiers on the label are correct but the label has the incorrect color coding.
- Root cause
- Mixed-up of materials/components
- Action
- Biomet 3i sent an Urgent - Medical Device Removal letter / and by phone call on and dated June 25, 2014, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. If you have any Certain BellaTek Encode Healing Abutment(s) of the affected lot in your practice, please return the abutment(s) to BIOMET 3i in exchange for Cettain BellaTek Encode Healing Abutments, IEHA458 with a proper color stripe. Alternatively, you may return the Certain BellaTek Healing Abutment(s) for full credit. We also ask that you confirm receipt of this notification by completing the attached Recall Return Response Form, indicating your preference for replacement or credit and/ or notifying us if the product was used. Thank you for your support - we are committed to maintaining high standards of quality for all our products and apologize for any inconvenience this may have caused you. If you have any further questions, please contact Customer Service at 1-800-342-5454 for assistance.
- Quantity
- 6 units
- Distribution
- US Distribution to NH and TX.
- Lot, serial or model codes
- Model Number IEHA458, Lot number: 1167545.
- 510(k) numbers
- ["K072642"]
- PMA numbers
- not on file
- Product code
- NHA
- Firm FEI
- 1038806
- Firm city
- Palm Beach Gardens
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28