Hospira Inc.: Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT FILL 900 ABOVE LINE 2) List Numbers 43423-01 and 43423-21; LIST NO. 43423; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; 1400. 3) List Numbers 43445-01 and 43445-11; LIST NO. 43445; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; DO NOT FILL 1900 ABOVE THIS LINE. 4) List Number 43496-01.
- Recall number
- Z-2593-2016
- Recalling firm
- Hospira Inc.
- Product
- Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT FILL 900 ABOVE LINE 2) List Numbers 43423-01 and 43423-21; LIST NO. 43423; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; 1400. 3) List Numbers 43445-01 and 43445-11; LIST NO. 43445; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; DO NOT FILL 1900 ABOVE THIS LINE. 4) List Number 43496-01.
- Status
- Terminated
- Initiated
- 2013-09-10
- Posted
- not on file
- Terminated
- 2016-12-15
- Reason
- Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.
- Root cause
- Process control
- Action
- URGENT MEDICAL DEVICE RECALL letters dated September 10, 2013 were distributed to direct accounts notifying them of the recall. The letter requests that customers to ensure that they only use liners and canisters of corresponding sizes (e.g. a 1000mL liner with a 1000mL canister) and to only use Hospira components. The letter further advises customers of an ongoing recall of the 1000mL Receptal systems. Customers were requested to complete and return the attached reply form and to notify other healthcare providers within their organization.Customers who further distributed the Receptal systems were requested to forward the letter to their customers and to request that the contact Stericycle at 1-888-943-5175 (Monday - Friday, 8:00 AM - 5:00 PM, CST) to obtain a reply form.
- Quantity
- 67,410,861 liners and 67,410,861 lids/canisters in total
- Distribution
- *** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa
- Lot, serial or model codes
- 43423-04-01 43423-04-11 43423-04-21 43445-04-01 43445-04-11 43449-04-01 43449-04-11 43496-04-01
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GCX
- Firm FEI
- 3013319212
- Firm city
- Lake Forest
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28