SenTec AG: Membrane Changer Single-Use, REF MC, each package contains 3 changers; Membrane Changer Reloadable, REF MC-R, each package contains 3 changers; and Membrane Changer Insert, REF MC-I, each package contains 5 inserts. All are components of the Sentec Digital Monitoring System (SDMS).
- Recall number
- Z-2592-2023
- Recalling firm
- SenTec AG
- Product
- Membrane Changer Single-Use, REF MC, each package contains 3 changers; Membrane Changer Reloadable, REF MC-R, each package contains 3 changers; and Membrane Changer Insert, REF MC-I, each package contains 5 inserts. All are components of the Sentec Digital Monitoring System (SDMS).
- Status
- Open, Classified
- Initiated
- 2023-07-25
- Posted
- 2023-09-15
- Terminated
- not on file
- Reason
- During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance.
- Root cause
- Process control
- Action
- The recalling firm notified Sentec distributors in a letter flagged Guidance for Sentec Distributors dated 7/25/2023 via email on 7/25/2023 (letter #RF-013993-a). The letter listed the devices impacted, the defect, how it manifests itself, and the risks (impact on measurement readings, impact on the sensor membrane, and harm to patient skin). Actions for the distributor and subdistributor level were provided. Subdistributors were to be informed who received the affected products by forwarding them this guidance document and related attachments using a read receipt on the email. The distributor is to follow-up with each subdistributor to ensure they understand the information provided and that each subdistributor will implement all actions as required and on time. The distributor is to confirm that these tasks have been executed by 7/31/2023. Potential return of product by the distributors will occur. All end customers within the distributor's territory are to be informed using the Customer Information Letter dated 7/25/2023 (letter #RF-013992-a) which was attached to the Guidance for Sentec Distributors. These letters were to be sent via email using a read receipt. The Customer Information Letter also contained information on the devices impacted, what the defect is, how it manifests itself, and the risks (impact on measurement readings, impact on the sensor membrane, and harm to patient skin). The end user is to identify any lots they may still have at their institution and if any of the affected products are in stock, the recalling firm recommends they follow the visual inspection steps shown in the enclosed Appendix I. If a defective membrane is detected, they are to replace it immediately. The customer is referred to Appendix I and II or the Instruction Manuals listed in the letter for the IFU procedure and picture of what a defective membrane may look like. Enclosed Appendix III contains a table listing the affected lot numbers for the Membran
- Quantity
- REVISED 9/21/2023-15,065 packages
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- REF MC: 230441, 230224, 230209, 230191, 230185, 230167, 230165, 230146, 230125, 230016, 225261, 225226, 225183, 225173, 225149, 225147, 225126, 225122, 225088, 225073, 225063, 225006; UDI-DI numbers 07640121880933, 17640121880930; REF MC-R: 230461, 230329, 230351, 230235, 230083, 230053, 230043, 230008, 230005, 225253, 225221; UDI-DI numbers 07640121880087, 17640121880084, 27640121880081; REF MC-I: 230372, 230339, 230336, 230229, 230206, 230132, 230117, 230051, 230044, 230012, 225299, 225298, 225280, 225255, 225218, 225216, 225190, 225185, 225184, 225175, 225120, 225091, 225089, 225064, 225058, 225033, 225023, 221064, 221056; UDI-DI numbers 07640121880506, 17640121880503. REVISED 9/21/2023: Additional lot number added for the U.S. for REF MC-I - 225119
- 510(k) numbers
- ["K041548","K101690","K151329"]
- PMA numbers
- not on file
- Product code
- LKD
- Firm FEI
- 3004149774
- Firm city
- Therwil
- Firm state
- not on file
- Firm country
- Switzerland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28