FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: MAYFIELD INFINITY XR2 BASE UNIT, Part Number A2079 The MAYFIELD Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.
- Recall number
- Z-2592-2018
- Recalling firm
- Integra LifeSciences Corp.
- Product
- MAYFIELD INFINITY XR2 BASE UNIT, Part Number A2079 The MAYFIELD Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.
- Status
- Terminated
- Initiated
- 2018-06-29
- Posted
- not on file
- Terminated
- 2020-02-26
- Reason
- There is a potential for the knob assembly in the swivel adapter to fracture/break during use.
- Root cause
- Device Design
- Action
- The firm, Integra, sent an "Urgent Voluntary Medical Device Correction Field Safety Notice" dated 6/29/18 to customers. The notification letter instructed customers to perform the following: 1) Review your MAYFIELD XR2 instructions for use. Please pay particular attention to the CAUTIONS listed on page 5 of the accompanying instructions for use section as noted below. " Always make sure the base unit is properly secured to the operating table " The base unit must not be used if the device appears to be damaged or functioning incorrectly " Over-tightening the mechanisms adjustment screws may result in damage to the unit " Over extending or overloading the base unit may result in unintended movement, shortened product life and/or damage to the unit " Do not alter the construction of this device as it may result in serious patient injury " Do not steam sterilize; components may be damaged by high heat leading to device damage and reduced performance 2) Visit Integra website and review video highlighting instructions for use at the following link: https://www.integralife.com/mayfield-imaging-cranial-stabilization-system/category/cranial-stabilization-mayfield-imaging-cranial-stabilization-system 3) Complete and return the attached Field Safety Notice Acknowledgement Form promptly via email: FCA@integralife.com or Fax to: 1-609-750-4220. Distributors were also instructed to: "We also kindly ask you to contact the final customers who may have the affected products and provide them with this letter." On July 20, 2018, an additional Urgent Voluntary Medical Device Recall letter was distributed to customers. The letters ask customers to perform the following: 1. Allow your sales representative to check the current knob assembly on your MAYFIELD Infinity XR2 Radiolucent Base Unit (A2079), Standard; MAYFIELD Infinity XR2 Low Profile Base Unit (A2079A); MAYFIELD Infinity XR2 Radiolucent Base Unit, Extended (A2079E) and MAYFIELD Spine Table Adaptor (
- Quantity
- 409 units
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV; and to countries of:: Australia, Canada, China, Dominican Republic, Hong Kong, India, Japan, Korea, Malaysia, Mexico, Pakistan, Panama, Peru, Singapore, Sri Lanka, Taiwan, and Thailand.
- Lot, serial or model codes
- 57218; 58242; 59524; 61083; 61084; 61533; 61559; 62801; 63826; 63949; 65966; 65967; 66950; 67912; 70432; 70564; 74838; 76382; 78005; 79436; 79475; 81218; 84107; 86162; 88507; 89760; 1468037; 1558488; 1664697; 1755770; 1485531; 1726814; 1737849; 1795116; 1939414; 2051032; 2135345; 2162789; 2247499; 2353536; 2380794; 2450908; 2463223; 2497401; 2517644; 2556854; 2642914; 2642922; 2740222; 2760809; 2760810; 2781235; 2781238
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GDJ
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28