Ion Beam Applications S.A.: Proteus 235
- Recall number
- Z-2592-2017
- Recalling firm
- Ion Beam Applications S.A.
- Product
- Proteus 235
- Status
- Terminated
- Initiated
- 2017-02-09
- Posted
- 2017-06-19
- Terminated
- 2023-07-14
- Reason
- It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.
- Root cause
- Device Design
- Action
- Ion Beam Application S.A. (IBA) will bring the defect into compliance: 1. Providing recommendations for user actions which, if followed, will prevent dark current. In order to prevent the error from occurring, the IBA customers are provided with a specific QA map that would detect dark current and instructions on how and when to use it (Annex 1). This guidance was included in IBAs User Notice which was sent to users on February 9, 2017. 2. IBA is developing a long term solution consisting of a new IC-cyclo (ionization chamber located at the cyclotron exit) and a Beam Absence Electronic Unit (BAEU) along with a software upgrade. This solution will detect presence of dark current between irradiations. It will prevent starting a new irradiation if dark current is present. This solution is currently in test phase and will be completed in two years. 3. Manufacturer will provide the FDA with progress reports on the effectiveness of the plan, including the number of electronic products repaired on a quarterly basis. The first progress report will be provided to the FDA by September 15, 2017 to the correspondence identified below. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2.Any future customer notification letters to be sent must comply with the requirements of 21 CFR 1003.21. This includes, but is not limited to, the prescribed packaging and the statement that you will remedy the defect or bring the product into compliance at no charge. For further questions please call (571) 449-4992.
- Quantity
- 18 systems ( 10 in the US)
- Distribution
- Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
- Lot, serial or model codes
- Serial Numbers: PAT.000,PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.111, PAT.112, PAT.113, PAT.114, SAT.115, SAT.116, SAT.117, SAT.119, SAT.120, SAT.122, SAT.123, SAT.126
- 510(k) numbers
- ["K152224"]
- PMA numbers
- not on file
- Product code
- LHN
- Firm FEI
- 3000256071
- Firm city
- Louvain La Neuve
- Firm state
- not on file
- Firm country
- Belgium
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28