Skip to the content

FDA Medical Device Recalls (openFDA)

ELITech Clinical Systems SAS: ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level). Model/Catalog Number: 55117

Recall number
Z-2591-2018
Recalling firm
ELITech Clinical Systems SAS
Product
ENVOY 500 ISE CALIBRATOR KIT, reference 55117 (contained 6 x 20 mL Calibrator High level and 6 x 20 mL Calibrator Low level). Model/Catalog Number: 55117
Status
Terminated
Initiated
2017-04-18
Posted
not on file
Terminated
2020-04-17
Reason
ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.
Root cause
Under Investigation by firm
Action
ELITech Clinical Systems SAS notified consignees on April 18, 2018 via USPS certified letter, dated April 10, 2018 titled Product Recall Notice. The Product Recall Notice provided customers/consignees reason for recall, actions to be taken, and included a confirmation of receipt of product recall notice form to be filled out and returned by customer/consignee.
Quantity
not on file
Distribution
U.S. only. No gov. accounts. No international.
Lot, serial or model codes
Lot number 7038 , Manufacture date 03/2017, Expiration date 02/2019
510(k) numbers
["K945271"]
PMA numbers
not on file
Product code
JIX
Firm FEI
3007662974
Firm city
Sees
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register