Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography

Recall number
Z-2591-2017
Recalling firm
Merge Healthcare, Inc.
Product
Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography
Status
Terminated
Initiated
2017-05-08
Posted
not on file
Terminated
2019-08-08
Reason
When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.
Root cause
Software design (manufacturing process)
Action
Merge sent an Urgent Medical Device Recall letter dated May 5, 2017, to all affected customers. Customers were instructed to be aware of the following workaround: Do not have the simple close study red x icon available in the reading physicians toolbars. This will force them to set status on close which also presents the warning Not all images have been reviewed, do you want to set status anyway and close study or continue reviewing unviewed images? YOUR RESPONSE TO THIS NOTIFICATION IS REQUIRED By replying to this notification, you will be taking the necessary steps to ensure that the issue listed in this document will be resolved. Please reply using the enclosed form and the return addressed envelope. Customers with questions were instructed to send an email to recall@merge.com. For questions regarding this recall call 877-741-6369.
Quantity
47 sites have the potentially affected versions
Distribution
Nationwide Distribution
Lot, serial or model codes
Versions 7.0.2.1, 7.0.2.2, 7.0.2.3, 7.0.2.3 P1, and 7.0.2.3 P2
510(k) numbers
["K082144"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register