FDA Medical Device Recalls (openFDA)
Miramar Labs, Inc.: MiraDry System consisting of the DTS3000 Console, MiraDry Handpiece (MD3001-HP), the miraDry biotip and the miraDry Treatment Template system. Miramar Labs, Inc. 445 Indio Way Sunnyvale, CA 94085 Intended use: Treatment of primary axillary hyperhidrosis.
- Recall number
- Z-2591-2011
- Recalling firm
- Miramar Labs, Inc.
- Product
- MiraDry System consisting of the DTS3000 Console, MiraDry Handpiece (MD3001-HP), the miraDry biotip and the miraDry Treatment Template system. Miramar Labs, Inc. 445 Indio Way Sunnyvale, CA 94085 Intended use: Treatment of primary axillary hyperhidrosis.
- Status
- Terminated
- Initiated
- 2011-04-24
- Posted
- 2011-06-17
- Terminated
- 2011-06-17
- Reason
- Reports of skin burns on axilla after treatment with the miraDry System.
- Root cause
- Device Design
- Action
- The firm,Miramar, has a service provider in Japan that assists in dealing with customers. The service providers were provided a copy of the investigation report and communicated directly with the customers to set up appointments for the field correction. The customers were personally visited by a representative from Miramars service provider in Japan. All devices had field correction performed on April 26-28, 2011.
- Quantity
- 8 units
- Distribution
- International only: Japan
- Lot, serial or model codes
- Serial numbers: 10H0158, 10H0160, 10H0161, 10H0167, 10H0172, 10H0185, 10H0193, 11H0035.
- 510(k) numbers
- ["K103014"]
- PMA numbers
- not on file
- Product code
- OUB
- Firm FEI
- 3008082710
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28