FDA Medical Device Recalls (openFDA)
BioFire Diagnostics, LLC: FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software
- Recall number
- Z-2588-2023
- Recalling firm
- BioFire Diagnostics, LLC
- Product
- FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software
- Status
- Open, Classified
- Initiated
- 2023-04-25
- Posted
- 2024-09-30
- Terminated
- not on file
- Reason
- The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel
- Root cause
- Software design
- Action
- On 4/25/23 correction notices were mailed to customers who were told to update the GI pouch module software. Instructions for installation can be downloaded here: https://www.biofiredx.com/e-labeling/ITIFA20GI21. This technical note and download cannot be opened via Firefox browser; please use a different browser. Once the pouch module software has been updated, complete and return the webform located at: https://www.biofiredx.com/gisoftware-5747 or return the acknowledgement form via email to FCA5747@biomerieux.com Questions or concerns can be directed to Technical Support at biofiresupport@biomerieux.com or via telephone by dialing 1.800.735.6544 and selecting option 5.
- Quantity
- 61,069
- Distribution
- US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW
- Lot, serial or model codes
- Model/UDI-DI: RFIT-ASY-0104/00815381020116, RFIT-ASY-0116/00815381020109 using software versions prior to FLM2-SFW-0040 v2.0.3.1/BFS0001-5539 v2.0.200.10
- 510(k) numbers
- ["K140407","K160459","K143005"]
- PMA numbers
- not on file
- Product code
- PCH
- Firm FEI
- 3002773840
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28