Skip to the content

FDA Medical Device Recalls (openFDA)

ARJOHUNTLEIGH POLSKA Sp. z.o.o.: Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds.

Recall number
Z-2588-2021
Recalling firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Product
Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds.
Status
Terminated
Initiated
2021-06-10
Posted
not on file
Terminated
2026-02-03
Reason
The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.
Root cause
Nonconforming Material/Component
Action
The Field Safety Notice dated 6/9/2021 was issued in two different mailings, one on 6/10/2021 and the second on 6/18/2021. The letter explained the issue and provided instructions for use until the product is corrected by Arjo personnel. All caregivers and users of the bed with IndiGo were to be made aware of the Field Safety Notice. A Customer Response Form was enclosed for completion to indicate the number of beds currently listed at the facility and that the customer has read and understands the Field Safety Notice. If the bed has been sold or moved, the customer is to provide information on the new facility.
Quantity
367 units
Distribution
US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA, and SD.
Lot, serial or model codes
Manufacturing dates 5/25/2018-2/15/2021 (printed on both the bed and IndiGo module if the module was assembled on the bed at the time of manufacture, or printed on the module if it was later installed on the bed.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FNL
Firm FEI
3007420694
Firm city
Komorniki
Firm state
not on file
Firm country
Poland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register