Deerfield Imaging, Inc.: Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.
- Recall number
- Z-2586-2021
- Recalling firm
- Deerfield Imaging, Inc.
- Product
- Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.
- Status
- Terminated
- Initiated
- 2021-08-25
- Posted
- not on file
- Terminated
- 2024-04-30
- Reason
- ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.
- Root cause
- Device Design
- Action
- A Recall notification letter titled, "URGENT FIELD CORRECTION NOTICE" was sent to consignees on 08/25/2021 via Fed-Ex. E-mail communications were also sent to the customers on 8/26/2021. The letter instructs the consignee to fill out the second page of the notice and return it to the firm. The letter states that the IMRIS Customer Service will contact the consignee to schedule implementation of this corrective action with the facility. The letter mentions that, "IMRIS recommends these preventive actions: Per the OR Table Operator Manuals, the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. During usage, verify the table position is locked by manually pushing the table to ensure it does not exhibit rotational movement. If the rotational lock is observed to not function properly, please contact IMRIS Customer Service immediately."
- Quantity
- 3 units
- Distribution
- Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR, UT, TN, NC and the countries of Canada, Australia, Sweden, Japan.
- Lot, serial or model codes
- UDI: 00857534006462; Model Number: 114093-600; Serial Number: 10007411, 10002996, 10004515.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KXJ
- Firm FEI
- 3010326005
- Firm city
- Minnetonka
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28