Skip to the content

FDA Medical Device Recalls (openFDA)

Telefelx Medical: WECK IMA/ENT Blade Electrode (1.9 cm), REF 809335, Rx Only Sterile, Teleflex Medical, RTP, NC The blade electrode is used in a wide variety of surgical procedures. The most common uses for electrodes with extended insulation is in cardiac bypass surgery (used in the take-down of the internal mammary artery or IMA) and ear, nose and throat surgery. They are sold non-sterile and are used one time and then disposed.

Recall number
Z-2584-2010
Recalling firm
Telefelx Medical
Product
WECK IMA/ENT Blade Electrode (1.9 cm), REF 809335, Rx Only Sterile, Teleflex Medical, RTP, NC The blade electrode is used in a wide variety of surgical procedures. The most common uses for electrodes with extended insulation is in cardiac bypass surgery (used in the take-down of the internal mammary artery or IMA) and ear, nose and throat surgery. They are sold non-sterile and are used one time and then disposed.
Status
Terminated
Initiated
2010-07-30
Posted
2010-09-27
Terminated
2011-04-21
Reason
Complaints were received regarding peeling and melting of the insulation of the cautery tip of the IMA/ENT blade electrodes.
Root cause
Other
Action
Teleflex Medical sent an "Urgent Medical Device Recall" letter dated July 30, 2010, to all customers. The letter described the product, problem and actions to be taken by customers. The customers were instructed to immediately discontinue use, quarantine any affected products and return all of the affected product to Teleflex Medical; forward the letter to their customers to retrieve product from those locations and complete and return the Recall Acknowledgement Form via fax at 1-866-804-9881. A second mailing to non-responding customers on the consignee list will be mailed approximately four to six seeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or as an alternative, they may be contacted by phone, e-mail, or fax. Note: A record of these contacts will be maintained. Affected products received by Teleflex Medical will be destroyed upon receipt. Records will be maintained accordingly. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
Quantity
190,344 eaches
Distribution
Worldwide distribution: USA and countries of Canada, Australia and Singapore.
Lot, serial or model codes
Catalog number: 809335, Lot numbers: 2218273, 2227760, T1213320, T1214910, 01K0700276, 01L0700066, 01M0700224, 01E0800244, 0G0800102, 01G0800132, 01J0800040, 01J0800288, 01J0800376, 01K0800195, 01K0800281, 01L0800228 and 01D0900449.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GEI
Firm FEI
3005747797
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register