Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA (HQ), Inc.: SynFix-LR Spinal Implant System (Implant Holder for SynFix-LR and surgical technique brochure) The implant holder is used to place the implant into the vertebral disc space.

Recall number
Z-2583-2010
Recalling firm
Synthes USA (HQ), Inc.
Product
SynFix-LR Spinal Implant System (Implant Holder for SynFix-LR and surgical technique brochure) The implant holder is used to place the implant into the vertebral disc space.
Status
Terminated
Initiated
2010-07-23
Posted
not on file
Terminated
2014-12-18
Reason
A potential for the implant holder tip to break off and lodge in the SynFix-LR implant, which pose a potential risk to the patient.
Root cause
Error in labeling
Action
Synthes Spine Hospital sent an ' URGENT: MEDICAL DEVICE LABELING CORRECTION" letter dated August 20, 2010 to all customers. The letter described the labeling issue and actions to be taken by the customers. The customers were instructed to review the supplementary labeling provided; contact their Synthes Spine Sales Consultant to arrange for provision of the Holding Forceps (p/n 388.401) prior to each SynFix-LR procedure; and complete and return the enclosed Verification Section form to Synthes via fax at 610-251-9005 or email: FieldAction@Synthes.com. If you have any questions, please call 1-800-620-7025 x 3500 or contact your Synthes Spine Sales Consultant.
Quantity
not on file
Distribution
Worldwide Distribution: USA and the countries of Canada, Japan, Switzerland, Australia, Mexico, Malaysia, New Zealand, and Singapore.
Lot, serial or model codes
Catolog Number: 03.802.039 - all lot numbers
510(k) numbers
["K053508"]
PMA numbers
not on file
Product code
MQP
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register