Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes (USA) Products LLC: TI Single Vector Distractor Body with Left Foot/20mm; Catalog ID 487.963 The Titanium Single Vector Distractor and Proximal Foot Plates are intended for mandibular bone lengthening where gradual bone distraction is required, including conditions such as congenital mandibular deficiencies or post-traumatic defects.

Recall number
Z-2582-2016
Recalling firm
Synthes (USA) Products LLC
Product
TI Single Vector Distractor Body with Left Foot/20mm; Catalog ID 487.963 The Titanium Single Vector Distractor and Proximal Foot Plates are intended for mandibular bone lengthening where gradual bone distraction is required, including conditions such as congenital mandibular deficiencies or post-traumatic defects.
Status
Terminated
Initiated
2016-05-23
Posted
not on file
Terminated
2016-12-13
Reason
DePuy Synthes is voluntarily initiating a recall due to a possibility the fastener on the Titanium Single Vector Distractor Body may become prematurely separated from the Proximal Foot Plate.
Root cause
Under Investigation by firm
Action
The firm, DePuy Synthes, mailed out an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated 5/23/2016 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to review your inventory and immediately take the following actions: If you DO have any of the identified devices call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number, complete the Verification Section (page 3 of the letter) indicating the number of devices that were found and return the Verification Section (page 3 of the letter) with the product to the following: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132, by fax 844-449-8498 or scan/email: Synthes8096@stericycle.com., and complete the Verification section even if you DO NOT have the identified product. Note: If the Verification Section is answered on behalf of more than one facility and/or individual clearly indicate the name and address of the facility and/or individual. If you have any questions, please call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
Quantity
182 units
Distribution
Worldwide Distribution-US (nationwide) distribution to states of: AZ, CA, CO, DC, FL, IL, MA, MI, NJ, NY, PA, TX, UT, WI, and WV; and countries of: Australia, Canada and Switzerland.
Lot, serial or model codes
Part #487.963, Lot # A4JX956; 4038561; 4091230; 4281408; 4294703; 4454536; 4475459; 5041111; 5041112; 5414450; 5414451; 5505099; 5505140; 5621104; 5621108; 5621109; 5621110; 6291848; 6879052; 7626782; 7865438; 9890200; mfg. 09JUN1999 -26APR2016, Exp. 22MAR2020-26APR2036
510(k) numbers
["K981075"]
PMA numbers
not on file
Product code
MQN
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register