FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: Discovery RT labeled as: a. MID BJG; b. HVY BJG
- Recall number
- Z-2581-2018
- Recalling firm
- GE Healthcare, LLC
- Product
- Discovery RT labeled as: a. MID BJG; b. HVY BJG
- Status
- Terminated
- Initiated
- 2018-05-24
- Posted
- not on file
- Terminated
- 2019-02-08
- Reason
- Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
- Root cause
- Component change control
- Action
- GE Healthcare notified customers on about 05/24/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter outlined the reason for recall and notified customers that a GE Healthcare representative would be contacting them to arrange for the correction of the devices. Any questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or a local Service Representative.
- Quantity
- 42 units total
- Distribution
- Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
- Lot, serial or model codes
- Model Number 2374682-17. a. MID BJG. Serial Number CBCIG1800017HM (System ID 847695CTRT, UDI 01008406821186991118020021CBCIG1800017HM), Serial Number CBCIG1700070HM (System ID 251665MCIRT, UDI 01008406821186991117120021CBCIG1700070HM), Serial Number CBCIG1700064HM (System ID 973831RT, UDI 01008406821186991117120021CBCIG1700064HM), Serial Number CBCIG1800001HM (System ID 708747RT590, UDI 01008406821186991118010021CBCIG1800001HM); b. HVY BJG. Serial Number CBCIG1700062HM (System ID 904819RTCT, UDI 01008406821186991117120021CBCIG1700062HM), Serial Number CBCIG1700040HM (System ID 843692GSOCT, UDI 01008406821186991117100021CBCIG1700040HM)
- 510(k) numbers
- ["K132410"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28