Skip to the content

FDA Medical Device Recalls (openFDA)

Keystone Industries: Henry Schein Hard Line Chairside Reline kit-Resilient Acrylic Denture Relining Material KI Item No.: 0921940HS Henry Schein Item No.: 570-0718

Recall number
Z-2580-2023
Recalling firm
Keystone Industries
Product
Henry Schein Hard Line Chairside Reline kit-Resilient Acrylic Denture Relining Material KI Item No.: 0921940HS Henry Schein Item No.: 570-0718
Status
Open, Classified
Initiated
2023-04-10
Posted
2023-09-12
Terminated
not on file
Reason
The Hard Line Chairside Reline Kit (570-0718) (0921940HS) was labeled as Soft Line Chairside Reline Kit (570-0717) (0921250HS) on the outer box. The contents of the Kit and the IFU enclosed with the kit were correctly labeled as Hard Line materials
Root cause
Packaging process control
Action
Keystone Industries notified the wholesaler on 4/10/23 via via email to initiate a warehouse alert and hold further shipments. A follow-up Letter was sent to the Wholesaler by email on April 12, 2023, with instructions for conducting the recall. The Wholesaler in turn sent a Recall Notification to all end user facilities on April 13, 2023. Letter states reason for recall, health risk and action to take: 1. Contact the appropriate warehouses identified above and issue an alert to stop all shipments of the affected product. 2. Have each warehouse check their inventory to determine if any of the affected product remains in their stock and segregate all affected units from inventory for return. 3. Implement appropriate control measures to quarantine any future shipments of the affected product that may be in transit to your warehouses. 4. Review distribution records and trace all affected product kits sent to customers. 5. Send Recall Notification on Henry Schein letterhead to all identified customers that received product, instructing them to return all affected products to your appropriate distribution center. Provide a copy of the Customer Recall Notification Letter template to Keystone Industries for submission to the FDA. 6. Quarantine all product returned from customers until the completion of the recall. Recall should end within 15 working days of issuing the Customer Recall Notifications. 7. Provide Keystone with a final report which records the amount of product returned from each Henry Schein distribution warehouse, amount originally shipped to each customer, number of customers, and number of products returned from each customer. 8. Upon completion of the recall, return all affected products to Keystone Industries by requesting Return Materials Authorization (RMA) number and scheduling product pickup with your appropriate sales representative. Complete the attached Acknowledgement of Medical Device Correction Form and return within 5 days of receiving this
Quantity
240 units
Distribution
US Nationwide distribution in the states of IL, PA, FL NV, TX.
Lot, serial or model codes
UDI: +H65857007171/$$325120723B058BP Lot Number: 23B058B
510(k) numbers
not on file
PMA numbers
not on file
Product code
EBI
Firm FEI
3010002722
Firm city
Gibbstown
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register