Randox Laboratories Ltd.: Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.

Recall number
Z-2575-2018
Recalling firm
Randox Laboratories Ltd.
Product
Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.
Status
Terminated
Initiated
2018-06-08
Posted
not on file
Terminated
2018-11-05
Reason
The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.
Root cause
Under Investigation by firm
Action
On June 8, 2018, the firm issued an Urgent Medical Device Recall letter to affected consignees. The letter stated that Randox has confirmed Liquid Cardiac Control CQ5053 Lot: 4245CK is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. Customers were instructed to do the following: " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Update kits with corrected IFU's excluding Troponin T values. " Complete and return the vigilance response section of this form to technical.services@randox.com within five working days.) In addition, distributors were instructed to send a copy of the letter to all affected customers and to those who need to be aware within their organization.
Quantity
1 kit (US), 602 kits (OUS)
Distribution
Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam
Lot, serial or model codes
Lot 4245CK.
510(k) numbers
["K110534"]
PMA numbers
not on file
Product code
JJY
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register