Beckman Coulter Inc.: UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

Recall number
Z-2574-2025
Recalling firm
Beckman Coulter Inc.
Product
UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322
Status
Open, Classified
Initiated
2025-08-13
Posted
2025-09-17
Terminated
not on file
Reason
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
Root cause
Component change control
Action
On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System (LIS), Middleware, or on affected instruments. Consult your Medical Director to determine if a retrospective review of results is warranted. Beckman Coulter recommends posting this letter on or near the affected analyzers Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. Acknowledge receipt of the notification.
Quantity
not on file
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.
Lot, serial or model codes
UDIs: 15099590672423, 15099590381646, 15099590687922/ Serial Numbers: AT38527, AU02008, AU10078, AV28630, AW13187, AW29123, BC18100, RBB03201, AW43276, AY06074, AY06092, AY41556, AY43605, AZ02030, AZ03041, AZ18245, AZ22329, AZ39513, AZ46681, BA03048, BA03050, BA03055, BA03060, BA07061, BA07062, BA13227, BA15245, BA27414, BB16181, BB29286, BC08036, BC18099, AZ25376, BB25262, BB25261, BA25393
510(k) numbers
["K193124"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register