Medline Industries Inc: NAMIC Manifolds. (1) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038301, UPN H965640383011, 3 Valve Manifold, Off Star Handles, Ports on Right, Rx only, Sterile; (2) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038303, UPN H965640383031, 3 Valve Manifold, On Star Handles, Ports on Right, Rx only, Sterile; (3) NAMIC Angiographic MANIFOLD High Pressure, REF 70037200, UPN H965700372001, 2 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (4) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037202, UPN H965700372021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile. (5) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037301, UPN H965700373011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (6) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037303, UPN H965700373031, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (7) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038200, UPN H965700382001, 2 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (8) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038202, UPN H965700382021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (9) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038301, UPN H965700383011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (10) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038303, UPN H96570038301, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (11) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039202, UPN H965700392021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (12) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039301, UPN H965700393011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; and (13) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039303, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile.
- Recall number
- Z-2574-2021
- Recalling firm
- Medline Industries Inc
- Product
- NAMIC Manifolds. (1) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038301, UPN H965640383011, 3 Valve Manifold, Off Star Handles, Ports on Right, Rx only, Sterile; (2) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038303, UPN H965640383031, 3 Valve Manifold, On Star Handles, Ports on Right, Rx only, Sterile; (3) NAMIC Angiographic MANIFOLD High Pressure, REF 70037200, UPN H965700372001, 2 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (4) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037202, UPN H965700372021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile. (5) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037301, UPN H965700373011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (6) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037303, UPN H965700373031, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (7) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038200, UPN H965700382001, 2 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (8) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038202, UPN H965700382021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (9) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038301, UPN H965700383011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (10) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038303, UPN H96570038301, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (11) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039202, UPN H965700392021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (12) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039301, UPN H965700393011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; and (13) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039303, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile.
- Status
- Terminated
- Initiated
- 2021-08-06
- Posted
- not on file
- Terminated
- 2025-08-25
- Reason
- Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions.
- Root cause
- Process design
- Action
- The recalling firm issued letters dated and mailed on 8/6/2021 notifying consignees of the issue and requesting they immediately check their stock for the affected product and destroy it.
- Quantity
- 816,420 manifolds and stopcocks
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Australia, Japan, Latvia, Lebanon, Malaysia, Netherlands, Pakistan, People's Republic of China, Panama, Singapore, South Africa, South Korea, Thailand, and United Kingdom.
- Lot, serial or model codes
- Expiration dates 12/31/2023-6/30/2024. (1) REF 64038301 - Lot numbers 0000006342, 0000008482, 0000016519; UPN H965640383011 - Lot number 5642413; (2) REF 64038303 - Lot numbers 0000010144, 0000014189, 0000016524; UPN H965640383031 - Lot number 5642416; (3) REF 70037200 - Lot numbers 0000002033, 0000008570, 0000016521; UPN H965700372001 - Lot numbers 5649219, 5659335; (4) REF 70037202 - Lot numbers 0000014365, 0000016522; UPN H965700372021 - Lot numbers 5642417, 5642565, 5642566, 5642567, 5642568, 5643626, 5647596; (5) REF 70037301 - Lot numbers 0000002024, 0000006320, 0000010129; UPN H965700373011 - Lot number 5662011; (6) REF 70037303 - Lot numbers 0000002051, 0000006046; UPN H965700373031 - Lot numbers 5655475, 5662957, 5664026; (7) REF 70038200 - Lot numbers 0000001974, 0000006032, 0000010140, 0000014125; UPN H965700382001 - Lot numbers 5645162, 5649221, 5662013; (8) REF 70038202 - Lot numbers 0000007917, 0000010149, 0000016523; UPN H965700382021 - Lot numbers 5642414, 5649222, 5657918, 5660843; (9) REF 70038301 - Lot numbers 0000006189, 0000010147; UPN H965700383011 - Lot numbers 5642415, 5648171, 5649224, 5652296; (10) REF 70038303 - Lot numbers 0000006240, 0000008481, 0000010133, 0000010142, 0000014338; UPN H96570038301 - Lot numbers 5649225, 5650868; (11) REF 70039202 - Lot number 0000006312; UPN H965700392021 - Lot number 5662961; (12) REF 70039301 - Lot number 0000006055; UPN H965700393011 - Lot numbers 5642420, 5643629, 5652297, 5658658, 5662016; and (13) REF 70039303 - Lot number 0000016625.
- 510(k) numbers
- ["K782095","K842829"]
- PMA numbers
- not on file
- Product code
- DTL
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28