Philips Medical Systems (Cleveland) Inc: Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques.

Recall number
Z-2570-2025
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques.
Status
Open, Classified
Initiated
2025-08-05
Posted
2025-09-16
Terminated
not on file
Reason
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Root cause
Software design
Action
On 08/05/2025, the firm sent via FedEx an "Urgent Medical Device Correction" Letter informing customers of a software issue affecting the Pinnacle3 Radiation Therapy Planning System performing Advanced Region of Interest (ROI) Expansion and Contraction that results in a potential shift of the intended ROI. The software assumes the patient is always in the Head First Supine (HFS) position when transposing the Advanced Expansion to alternate positions. Customers are instructed to: -When using the Advanced ROI Expansion and Contraction feature in Pinnacle3, only use the Head First Supine (HFS) patient orientation. -Ensure that relevant clinical staff, including physicians, physicists, and dosimetrists, are informed about the potential for directional errors in ROI expansion/contraction under certain orientation conditions. -Customers may continue to use your system(s) in accordance with the intended use and by following the recommendations above. -Circulate this notice to all users of this device so that they are aware of the potential issue. -Retain this letter with their system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. -Complete and return the updated attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: Philips.recall@philips.com. Note: If customers have returned a response form already in reply to the initial customer letter, complete and sign this updated response form and return to Philips. Philips will provide a label to be affixed to the device that provides instruction to use the HFS position only. A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to attach /affix the label to the workstation (Reference 2025-EI-TPS-002). Questions-contact your local Philips representative: For North America, contact Philips Customer Care Center at 1-800-722-9377.
Quantity
810 (350 U.S. and 460 O.U.S.)
Distribution
U.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
Lot, serial or model codes
Pinnacle¿ Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model Number: 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Lot Number (Software Version): 18.0.5 Pinnacle TumorLOC 16.2 Model Number: 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Lot Number (Software Version): 16.2 Pinnacle TumorLOC 16.2.1 Model Number: 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Lot Number (Software Version): 16.2.1 Pinnacle TumorLOC 16.0.2 Model Number: 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Lot Number (Software Version): 16.0.2 Pinnacle TumorLOC 14.0 Model Number: 870200 UDI code: N/A Lot Number (Software Version): 14.0
510(k) numbers
["K041577","K090808","K170086"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3018004899
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register