FDA Medical Device Recalls (openFDA)
Abiomed, Inc.: Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 2. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 3. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 4. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L.
- Recall number
- Z-2568-2025
- Recalling firm
- Abiomed, Inc.
- Product
- Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 2. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 3. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 4. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L.
- Status
- Open, Classified
- Initiated
- 2025-08-20
- Posted
- 2025-09-19
- Terminated
- not on file
- Reason
- Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
- Root cause
- Device Design
- Action
- Abiomed notified consignees via letter on about 08/20/2025. Consignees were instructed to quarantine and cease use of all affected units, contact Abiomed Field Service team to initiate the remediation process, complete and return the business reply form with the assistance of your Abiomed Representative, forward the recall notification to any applicable personnel and notify customers if affected units were further distributed or transferred. Abiomed also posted a version of the recall notification letter on their website.
- Quantity
- 71 units
- Distribution
- US distribution to AR, AZ, CA, CO, CT, FL, GA, ID, MA, MI, MO, MS, NJ, NY, OH, PA, TX. International distribution to Canada, Germany, Great Britian, Italy, Netherlands, Norway.
- Lot, serial or model codes
- 1. Product Code: 0042-0000-US; GTIN: 813502010022; Serial Numbers: IC1026, IC1029, IC1043, IC1046, IC1040, IC1047, IC1088, IC1089, IC1093, IC1110, IC1167, IC1221, IC1225, IC1222, IC1258, IC1250, IC1272, IC1293, IC1320, IC1317, IC1353, IC1354, IC1355, IC1375, IC1377, IC1403, IC1446, IC1450, IC1452, IC1461, IC1496, IC1501, IC1508, IC1515, IC1516, IC1517, IC1518, IC1530, IC1534, IC1535, IC1539, IC1538, IC1531, IC1533, IC1540, IC1532, IC1551, IC1552, IC1550, IC1553, IC1646, IC1756, IC2255, IC2422, IC2489, IC3013. 2. Product Code: 0042-0000-US-L; GTIN: 813502010022; Serial Number: IC1281. 3. Product Code: 0042-0000-CA; GTIN: 813502011272; Serial Numbers: IC1543, IC1544. 4. Product Code: 0042-0000-EU; GTIN: 813502011289; Serial Numbers: IC1350, IC1383, IC1546, IC1548, IC1547, IC1549, IC1775, IC2243, IC4002. 5. Product Code: 0042-0000-UK; GTIN: 813502011296; Serial Numbers: IC1519, IC1542. 6. Product Code: 0042-0000-UK-L; GTIN: 813502011296; Serial Number: IC1493.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OZD
- Firm FEI
- 1220648
- Firm city
- Danvers
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28