FDA Medical Device Recalls (openFDA)
Natus Neurology Inc: Nicolet Ambulatory EEG
- Recall number
- Z-2568-2018
- Recalling firm
- Natus Neurology Inc
- Product
- Nicolet Ambulatory EEG
- Status
- Terminated
- Initiated
- 2018-02-01
- Posted
- not on file
- Terminated
- 2022-10-21
- Reason
- Potential to be able to import an ambulatory exam for a patient for which the exam was not started.
- Root cause
- Software design
- Action
- Customers were notified via letter on about 02/01/2018. Instructions to ensure the empty exam is removed appropriately after starting the ambulatory recording included: 1) Only shutdown Recorder via the Recorder UI, either from the File menu or the X button on the top right of the application; 2) If any of the four conditions above occurs and the 2.60 MB exam is not removed, the empty exam can be deleted from NicVue prior to importing the ambulatory exam, doing so will allow the patient ID check to occur normally. Customers were also instructed to provide the recall communication to all applicable users and complete and return the verification form.
- Quantity
- 171 units
- Distribution
- Distributed domestically to AR, AZ, CA, CO, DC, FL, GA, ID, IL, IN, MA, ME, MI, MO, N Y, NC, ND, NJ, NV, NY, PA, TN, TX, VA, WA, WI. Distributed internationally to Australia, Belgium, China, Germany, Hong Kong, Israel, Italy, Jordan, Kuwait, Russia, Slovenia, Tunisia, United Arab Emirates, Vietnam.
- Lot, serial or model codes
- UDI 00382830043670; Serial Numbers MBR120693, MBR100549, MBR100566, MBR120699, MBR090299, MBR090317, MBR150907, MBR100541, MBR120700, MBR120681, MBR120686, MBR120697, MBR130793, MBR110634, MBR150920, MBR150921, MBR160934, MBR120665, MBR110629, MBR090308, MBR150885, MBR090327, MBR120720, MBR130784, MBR130785, MBR120696, MBR100579, MBR090318, MBR150928, MBR100610, MBR110642, MBR090313, MBR100578, MBR100564, MBR100626, MBR120683, MBR100594, MBR090324, MBR090326, MBR090329, MBR100583, MBR100602, MBR100603, MBR100624, MBR120659, MBR120674, MBR100567, MBR140853, MBR100565, MBR110645, LBR06094, LBR06095, LBR06101, MBR120660, MBR100550, MBR120671, MBR150886, MBR150890, MBR150893, MBR150894, MBR100573, MBR150896, MBR150897, MBR150902, MBR150904, MBR150911, MBR150916, MBR150918, MBR150919, MBR150908, MBR150930, MBR150931, MBR100552, MBR100596, MBR100601, LBR06133, MBR110638, MBR140836, MBR100572, MBR110649, MBR100532, MBR100548, MBR110630, MBR120734, MBR150898, MBR100592, MBR090303, MBR090306, MBR100586, MBR100600, MBR150887, MBR150888, MBR100545, MBR100555, MBR120658, MBR120680, MBR120687, MBR160952, MBR140857, MBR120762, MBR100580, MBR100591, MBR140862, MBR150922, MBR140856, MBR140865, MBR150903, MBR140861, MBR140834, MBR140858, 140859, MBR140863, MBR140864, MBR140866, MBR150884, MBR150917, MBR160936, MBR160941, MBR160962, MBR160969, 80218, LBR120757, LBR120759, LBR120771, MBR110635, MBR110640, MBR110643, MBR110644, MBR110647, MBR120661, MBR120662, MBR120675, MBR120725, MBR120733, MBR140833, MBR140835, MBR140838, MBR100625, MBR150899, MBR110646, MBR150913, MBR090321, MBR090330, MBR100585, MBR100617, MBR110641, MBR120679, MBR120710, MBR120715, MBR120718, MBR120754, MBR120760, MBR120763, MBR130794, MBR140815, MBR150926, LBR06028, LBR06051, MBR100614, MBR100621, MBRL20672, MBR140811, MBR150929, 150923, MBR120653, MBR120713, MBR150925, MBR150891
- 510(k) numbers
- ["K021185"]
- PMA numbers
- not on file
- Product code
- OLV
- Firm FEI
- 3010611950
- Firm city
- Middleton
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28