FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Medline medical convenience kits labeled as: 1. NEWBORN RESUSCITATION KIT, Kit Number ACC010381C; 2. DRAWER 5 - POSITION 3, Kit Number ACC010532; 3. DRAWER #4C ADULT, Kit Number ACC010671; 4. DRAWER #2, Kit Number ACC010728.
- Recall number
- Z-2567-2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline medical convenience kits labeled as: 1. NEWBORN RESUSCITATION KIT, Kit Number ACC010381C; 2. DRAWER 5 - POSITION 3, Kit Number ACC010532; 3. DRAWER #4C ADULT, Kit Number ACC010671; 4. DRAWER #2, Kit Number ACC010728.
- Status
- Open, Classified
- Initiated
- 2025-07-25
- Posted
- 2025-09-19
- Terminated
- not on file
- Reason
- Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
- Root cause
- Component design/selection
- Action
- Medline Industries issued a MEDICAL DEVICE RECALL notice to it consignees on 07/25/2025 via email and USPS first class mail. The notice explained the issue with the component within the kit, potential risk to the patient, and requested the following: REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-25-158-FGX1 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 596 units
- Distribution
- US distribution to: AL, CA, DE, OR, TN, TX
- Lot, serial or model codes
- 1. ACC010381C: UDI/DI 10193489922127 (EA), 40193489922128 (CS), Kit Lot Numbers: 2. ACC010532: UDI/DI 10193489902358 (EA), 40193489902359 (CS), Kit Lot Numbers: 3. ACC010671: UDI/DI 10195327351090 (EA), 40195327351091 (CS), Kit Lot Numbers: 4. ACC010728: UDI/DI 10195327506544 (EA), 40195327506545 (CS), Kit Lot Numbers:
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28