Computerized Medical Systems Inc: Focal Radiation Treatment Planning Workstation. Focal Release 4.51.00 and above. Allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data form a Windows-based PC and transfer the information back to the RTP system. It is intended for use in radiation treatment planning using generally accepted contouring methods.

Recall number
Z-2562-2011
Recalling firm
Computerized Medical Systems Inc
Product
Focal Radiation Treatment Planning Workstation. Focal Release 4.51.00 and above. Allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data form a Windows-based PC and transfer the information back to the RTP system. It is intended for use in radiation treatment planning using generally accepted contouring methods.
Status
Terminated
Initiated
2010-11-09
Posted
2011-06-14
Terminated
2014-04-02
Reason
Focal: If a plan involving Dynamic MLC Delivery is sent from XiO to Focal and called up there in Plan Review, the angle of the gantry is set to zero for all beams. If this plan is then sent from Focal to the R&V System, the gantry angles all remain at zero which could lead to patient mistreatment.
Root cause
Software design
Action
Elekta CMS Software sent a USER NOTICE dated November 5, 2010, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. A return postcard was included for the customer to confirm receipt of the User Notice. The firm intends to correct the software so it does not modify anything about the plan sent from XiO. Customers were instructed to contact their local customer support office with any questions.
Quantity
155
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Algeria, Australia, Belarus, Belgium, Brazil, Canada, Chile, China, Egypt, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Latvia, Libya, Lithuania, Mexico, Morocco, Netherlands, New Zealand, Philippines, Poland, Romania, Singapore, Slovenia, South Africa, Spain, Suriname, Switzerland, Taiwan, Turkey, Turkmenistan, and the United Kingdom.
Lot, serial or model codes
Release 4.51.00 and higher
510(k) numbers
["K013112"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
1937649
Firm city
Maryland Heights
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register