FDA Medical Device Recalls (openFDA)
Merit Medical Systems, Inc.: Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF:K04-00791A An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.
- Recall number
- Z-2561-2024
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF:K04-00791A An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.
- Status
- Open, Classified
- Initiated
- 2024-06-11
- Posted
- 2024-08-12
- Terminated
- not on file
- Reason
- Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
- Root cause
- No Marketing Application
- Action
- On 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that appropriate personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities. 4. Follow the attached instructions for placing the recall labels on the products containing Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled. 5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification. 6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products. 7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers. For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
- Quantity
- 1,390 kits
- Distribution
- U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
- Lot, serial or model codes
- Catalog Number: K04-00791A UDI-DI code: 00884450148471 Lot Numbers: H2387533 H2408714 H2423858 H2447173 H2484854 H2509065 H2547204 H2557351 H2640509 H2669424 H2704230 H2770492 H2780739 H2792326 H2864128 H2869756 H2916432
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJA
- Firm FEI
- 1721504
- Firm city
- South Jordan
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28