FDA Medical Device Recalls (openFDA)

Nanosphere, Inc.: Verigene CYP2C19 Nucleic Acid Test (CYP2C19 Test). Each CYP2C19 kit consists of 20 Verigene CYP2C19 Nucleic Acid Test Cartridges and 20 Verigene CYP2C19 Nucleic Acid Extraction Trays (with Tip Holder Assemblies). Clinical laboratory test.

Recall number
Z-2559-2014
Recalling firm
Nanosphere, Inc.
Product
Verigene CYP2C19 Nucleic Acid Test (CYP2C19 Test). Each CYP2C19 kit consists of 20 Verigene CYP2C19 Nucleic Acid Test Cartridges and 20 Verigene CYP2C19 Nucleic Acid Extraction Trays (with Tip Holder Assemblies). Clinical laboratory test.
Status
Terminated
Initiated
2014-03-18
Posted
2014-08-28
Terminated
2015-07-02
Reason
Nanosphere has recieved reports related to an unacceptable increase in initial no-call rate when performing the Verigene CYP2C19 Nucleic Acid Test (2C19), which has been confirmed through internal testing.
Root cause
Under Investigation by firm
Action
Nanosphere sent an Field Notification Letter dated March 18, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letters included instructions for customers to: 1) discontinue use of the affected products, 2) discard the affected product, and 3) complete and return the attached for to obtain free-of-charge replacement kits. Customers with any questions about this recall should contact their Nanosphere technical support representative at 1 888 837-4436. .
Quantity
78 kits
Distribution
Worldwide Distribution - US including the states of AL, CA, MD, ND, NJ, OK and TX., and internationally to Bulgaria, China and New Zealand.
Lot, serial or model codes
(Catalog No. 20-005-019)    Kit Numbers (US Distribution): 051013019-9, 6288-1, 6547-1, 6420-1, 5478-1, 5802-1, 6190-1, 7360-3, 5452-2, 6499-1, 6414-1, 7183-1, 7511-1, 5225-1, 5225-2, 6495-1, 5760-1, 5779-1, 7259-1, 6783-1, 7435-1    Kit Numbers (Foreign Distribution): 7300-4, 5500-1, 5908-3, 5970-3, 6402-1, 6402-2, 6105-1, 6123-1, 6793-1, 6827-1, 6224-1
510(k) numbers
["K120466"]
PMA numbers
not on file
Product code
NTI
Firm FEI
3006028115
Firm city
Northbrook
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register