Skip to the content

FDA Medical Device Recalls (openFDA)

Vital Scientific N.V.: V-Twin, Model # 6002-800, UDI: 03661540600180 Product Usage: An in-vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.

Recall number
Z-2558-2019
Recalling firm
Vital Scientific N.V.
Product
V-Twin, Model # 6002-800, UDI: 03661540600180 Product Usage: An in-vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.
Status
Open, Classified
Initiated
2016-01-18
Posted
not on file
Terminated
not on file
Reason
Instrument stopped working due to a software lockup, and no patient results are produced.
Root cause
Software Manufacturing/Software Deployment
Action
In very close cooperation with the consignees, the affected end users running on EPROM version V1.2.3 were contacted. Initially, a temporary work around (reset of the software) was communicated with the consignees. The consignees devices were installed to document all errors during runtime .In addition, a new software was developed and tested. The new software resulted in version 1.2.5 of EPROM, which was released June 23, 2016. Consignees of the V-Twin all received a field update kit with the new software.
Quantity
2
Distribution
US in the state of New York.
Lot, serial or model codes
Serial #s: 13-3341 and 13-3344
510(k) numbers
["K973628"]
PMA numbers
not on file
Product code
JJF
Firm FEI
2708
Firm city
Dieren
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register