Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Accolade TMZF Plus Hip Stem: Catalog # 3020-0537 Accolate d TMZF Hip Stem # 5 Catalog #6020-3535 Accolade 132 Size 3.5 Catalog # 6021-0230 Accolade Plus TMZF Hip Stem #2 Catalog # 6021-0335 Accolade Plus TMZF Hip Stem #3 Catalog #3021-0435 Accolade TMZF Hip Stem #4 Catalog # 6021-0435 Accolade Plus TMZF Hip Stem #4 Catalog #6021-2530 Accolade (127 deg) Size 2.5 Accolade (127 deg) Size 2.5 Catalog # 6021-4535 Accolade (127 DEG) Size 4.5 K994366; k023105; K032300 Manufactured by " Stryker Ireland, Carrigtwohill Industrial Estate Carrington County Cork, Ireland The subject hip stem is a single-use device intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty.

Recall number
Z-2557-2011
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Accolade TMZF Plus Hip Stem: Catalog # 3020-0537 Accolate d TMZF Hip Stem # 5 Catalog #6020-3535 Accolade 132 Size 3.5 Catalog # 6021-0230 Accolade Plus TMZF Hip Stem #2 Catalog # 6021-0335 Accolade Plus TMZF Hip Stem #3 Catalog #3021-0435 Accolade TMZF Hip Stem #4 Catalog # 6021-0435 Accolade Plus TMZF Hip Stem #4 Catalog #6021-2530 Accolade (127 deg) Size 2.5 Accolade (127 deg) Size 2.5 Catalog # 6021-4535 Accolade (127 DEG) Size 4.5 K994366; k023105; K032300 Manufactured by " Stryker Ireland, Carrigtwohill Industrial Estate Carrington County Cork, Ireland The subject hip stem is a single-use device intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty.
Status
Terminated
Initiated
2011-03-01
Posted
2011-06-13
Terminated
2012-10-10
Reason
Accolade TMZF Plus Hip Stems are being recalled due to grit blast media which was observed in the drive hole.
Root cause
Employee error
Action
Stryker Orthopaedics sent an Urgent Product Recall letter dated March 1, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to fax back the attached Product Recall Acknowledgement Form within 5 days to 201-831-6069. Customers were also instructed to: Contact the hospitals in their territory that have the affected product to arrange return of the product. Return the affected product via and Inter-Org Transfer to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, New Jersey 07430. Attach the fluorescent orange PRODUCT REMEDIATION sticker to their return, indicating the Product Remediation # RA 2009-459. Mark the outer box with the words "Product Recall." For any questions call (201) 972-2100.
Quantity
total of all units is 44
Distribution
Nationwide Distribution
Lot, serial or model codes
Catalog @ 3020-0537 Accolated TMZF Hip Stem # 5 Lot #31390203 8 units CAtalog #6020-3535 Accolade 132 Size 3.5 Lot #31446602R 0 units  CAtalog # 6021-0230 Accolade Plus TMZF Hip Stem #2 Lot #31515703 10 units Catalog # 6021-0335 Accolade Plus TMZF Hipe Stem #3 Lot 31586402 10 units Catalog #3021-0435 Accolade TMZF Hipe STem #4 31597205 1 unit CAtalog # 6021-0435 Accolade Plus TMZF Hipe STem #4 Lot #31597206 4 units Catalog #6021-2530 Accolade (127 deg) Size 2.5 Accolade (127 deg) Size 2.5 Lot #31486505 10 units Catalog # 6021-4535 Accolade (127 DEG) Size 4.5 Lot #31850001 1 unit
510(k) numbers
["K032300","K994366"]
PMA numbers
not on file
Product code
ISL
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register