IPG Medical Corporation: Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
- Recall number
- Z-2554-2026
- Recalling firm
- IPG Medical Corporation
- Product
- Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
- Status
- Open, Classified
- Initiated
- 2026-05-05
- Posted
- 2026-06-24
- Terminated
- not on file
- Reason
- Software anomaly that causes a false display of error code 5018.
- Root cause
- Device Design
- Action
- An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or koconnell@ipgmedical.com.
- Quantity
- 198 units
- Distribution
- US Nationwide distribution in the state of MA.
- Lot, serial or model codes
- Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051, MDUF250052, MDUF250053, MDUF250054, MDUF250055, MDUF250056, MDUF250057, MDUF250058, MDUF250059, MDUF250062, MDUF250063, MDUF250064, MDUF250065, MDUF250066, MDUF250070, MDUF250072, MDUF250073, MDUF250074, MDUF250075, MDUF250084, MDUF250085, MDUF250088, MDUF250089, MDUF250090, MDUF250093, MDUF250102, MDUF250103, MDUF250105, MDUF250132, MDUF250145, MDUF250158, MDUF250161, MDUF250162, MDUF250198, MDUF250201, MDUF250206, MDUF250207, MDUF250220, MDUF250237, MDUF250238, MDUF250239, MDUF250243, MDUF250245, MDUF250279, MDUF250280, MDUF250295, MDUF250296, MDUF250298, MDUF250310, MDUF250311, MDUF250313, MDUF250323, MDUF250344, MDUF250348, MDUF250349, MDUF250350, MDUF250353, MDUF250357, MDUF250364, MDUF250365, MDUF250369, MDUF250371, MDUF250372, MDUF250378, MDUF250379, MDUF250381, MDUF250382, MDUF250383, MDUF250386, MDUF250393, MDUF250396, MDUF250397, MDUF250398, MDUF250407, MDUF250408, MDUF250411, MDUF250412, MDUF250428, MDUF250432, MDUF260016, MDUF260018, MDUF260020, MDUF260031, MDUF260041, MDUF260044, MDUF260069, MDUF260071, MDUF260078, MDUF260088, MDUF260089, MDUF260090, MDUF260091, MDUF260092, MDUF260093, MDUF250087, MDUF250208, MDUF250244, MDUF250324, MDUF250326, MDUF250354, MDUF250356, MDUF250394, MDUF260042, MDUF260045, MDUF260076; Software Version 2.2.
- 510(k) numbers
- ["K232568"]
- PMA numbers
- not on file
- Product code
- GEX
- Firm FEI
- 3020221089
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28