FDA Medical Device Recalls (openFDA)
SEASPINE ORTHOPEDICS CORPORATION: Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System
- Recall number
- Z-2554-2025
- Recalling firm
- SEASPINE ORTHOPEDICS CORPORATION
- Product
- Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System
- Status
- Open, Classified
- Initiated
- 2025-08-06
- Posted
- 2025-09-11
- Terminated
- not on file
- Reason
- Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking.
- Root cause
- Device Design
- Action
- On 8/6/2025, recall notices were emailed to customers and distributors who were informed of the following: A laser marking error was identified in which the TA2-001003 inserters were incorrectly marked with the UDI-DI for model TA2-001002 (10889981290733). Updated Revision E Inserters have been shipped to you via overnight delivery. Kindly replace all existing Revision C/R1/R2 Inserters with the enclosed Revision E units. Return the Revision C Inserters to the recalling firm using provided RGA and prepaid shipping label. This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. Should you have any questions regarding this matter, please contact the firm at: dominiquepetach@orthofix.com
- Quantity
- 74
- Distribution
- Worldwide - US Nationwide distribution in the states of NV, TX, ID, MI, OH, SD, TN, PA, CT, CO, FL, WA, NC, OK, MN and the country of Hong Kong.
- Lot, serial or model codes
- UDI-DI: 10889981290733, 10889981437800. Lots: BT107376C, BT112289C, BT112779C, BT107376CR1, BT107376CR2, BT107376CR1R2, BT112289CR2, BT112779CR2
- 510(k) numbers
- ["K192132"]
- PMA numbers
- not on file
- Product code
- MAX
- Firm FEI
- 3012120772
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28