Skip to the content

FDA Medical Device Recalls (openFDA)

SEASPINE ORTHOPEDICS CORPORATION: Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System

Recall number
Z-2554-2025
Recalling firm
SEASPINE ORTHOPEDICS CORPORATION
Product
Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System
Status
Open, Classified
Initiated
2025-08-06
Posted
2025-09-11
Terminated
not on file
Reason
Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking.
Root cause
Device Design
Action
On 8/6/2025, recall notices were emailed to customers and distributors who were informed of the following: A laser marking error was identified in which the TA2-001003 inserters were incorrectly marked with the UDI-DI for model TA2-001002 (10889981290733). Updated Revision E Inserters have been shipped to you via overnight delivery. Kindly replace all existing Revision C/R1/R2 Inserters with the enclosed Revision E units. Return the Revision C Inserters to the recalling firm using provided RGA and prepaid shipping label. This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. Should you have any questions regarding this matter, please contact the firm at: dominiquepetach@orthofix.com
Quantity
74
Distribution
Worldwide - US Nationwide distribution in the states of NV, TX, ID, MI, OH, SD, TN, PA, CT, CO, FL, WA, NC, OK, MN and the country of Hong Kong.
Lot, serial or model codes
UDI-DI: 10889981290733, 10889981437800. Lots: BT107376C, BT112289C, BT112779C, BT107376CR1, BT107376CR2, BT107376CR1R2, BT112289CR2, BT112779CR2
510(k) numbers
["K192132"]
PMA numbers
not on file
Product code
MAX
Firm FEI
3012120772
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.