Skip to the content

FDA Medical Device Recalls (openFDA)

Bausch & Lomb Surgical, Inc.: enVista one-piece hydrophobic acrylic toric intraocular lens, Bausch and Lomb - Product Usage: is indicated for primary implantation in the capsular bag of the eye in adult patients for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision.

Recall number
Z-2552-2020
Recalling firm
Bausch & Lomb Surgical, Inc.
Product
enVista one-piece hydrophobic acrylic toric intraocular lens, Bausch and Lomb - Product Usage: is indicated for primary implantation in the capsular bag of the eye in adult patients for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision.
Status
Terminated
Initiated
2020-06-09
Posted
not on file
Terminated
2021-02-11
Reason
The toric axis marks were misaligned (from the expected position in relation to the haptics) on an implanted lens.
Root cause
Unknown/Undetermined by firm
Action
On 06/09/2020, BAUSCH and LOMB's Sales Representatives contacted customers by telephone and email. In addition, the firm sent a letter to customers dated, June 17, 2020 and titled, "MEDICAL DEVICE VOLUNTARY PRODUCT RECALL". The letter informs customers about a voluntary recall of one lot of enhanced enVista MX60ET intraocular lenses (IOLs). Customers should quarantine product and return product to the firm. Customers should complete the enclosed Medical Device Voluntary Recall Acknowledgement Form and contact the firm to obtain a Return Material Authorization Number and arrange for pickup of the identified product. The contact information for Bausch and Lomb Surgical Customer Service is 1-800-338-2020, Customer Service (option 2).
Quantity
5
Distribution
U.S. Nationwide distribution including in the states of IN, SC, CT, NV, and NC.
Lot, serial or model codes
Model number: MX60ET, Catalog number: MXUET  Lot numbers: Lot Number 3043212   Serial Numbers: 3043212002 (UDI (01)10757770555199(17)230131(21)3043212002); 3043212005 (UDI (01)10757770555199(17)230131(21)3043212005); 3043212006 (UDI (01)10757770555199(17)230131(21)3043212006); 3043212007 (UDI (01)10757770555199(17)230131(21)3043212007); 3043212008 (UDI (01)10757770555199(17)230131(21)3043212008)
510(k) numbers
not on file
PMA numbers
["P910056"]
Product code
HQL
Firm FEI
1119279
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register