Skip to the content

FDA Medical Device Recalls (openFDA)

Access Scientific LLC: POWERWAND XL 4Fr 8cm Maximum Barrier Kit (Introducer with an Extended Dwell Catheter made of ChronoFlex C with BioGUARD Technology), Catalog No. 94108, Device UDI 00859821006773/Pacakge UDI 10859821006770 - Product Usage: Introducer with Extended Dwell Catheter is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may also be used for power injection of contrast media up to a rate 8ml/sec, at a maximum of 325 psi fluid pressure.

Recall number
Z-2550-2020
Recalling firm
Access Scientific LLC
Product
POWERWAND XL 4Fr 8cm Maximum Barrier Kit (Introducer with an Extended Dwell Catheter made of ChronoFlex C with BioGUARD Technology), Catalog No. 94108, Device UDI 00859821006773/Pacakge UDI 10859821006770 - Product Usage: Introducer with Extended Dwell Catheter is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may also be used for power injection of contrast media up to a rate 8ml/sec, at a maximum of 325 psi fluid pressure.
Status
Terminated
Initiated
2020-06-15
Posted
not on file
Terminated
2024-10-30
Reason
Convenience kit mislabeled with the incorrect catheter effective length on the device label and the packaging label. Use may lead to a delay or interruption of treatment and user dissatisfaction or annoyance.
Root cause
Labeling Change Control
Action
On 06/15/2020, the firm notified affected customers via email and/or overnight mail of "URGENT: MEDICAL DEVICE RECALL" letter indicating the following: Immediately examine your inventory and quarantine product subject to correction. In addition, if you may have further distributed this product, please identify your customers and notify them a once of this product correction. Your notification to you customers may be enhanced by including a copy of this notification letter. Please take the following actions: 1. Review your inventory for the catalog and lot numbers specified in this notification and quarantine the product as appropriate. 2. You may continue to use the product specified in this notification with the understanding of the errors detailed in this notification, OR 3. Return the product specified in this notification for replacement product. 4. Complete and return the enclosed Recall Response Form as soon as possible. 5. If you choose to return the product, the firm will provide a prepaid shipping label for the return of affected product upon receipt of the Recall Response Form. 6. Report any adverse health consequences experienced with the use of this product to the firm. Events may also be reported to the Food and Drug Admirations MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you have any questions regarding this recall, please contact the firm's Senior Manager, Quality and Regulatory Affairs, via email or call 858-259-8333.
Quantity
1710
Distribution
US Nationwide distribution including in the state of NY.
Lot, serial or model codes
Lot number 269939, 270552, 270874, and 272261.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OWL
Firm FEI
3009549229
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register