Zimmer, Inc.: Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE

Recall number
Z-2550-2014
Recalling firm
Zimmer, Inc.
Product
Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE
Status
Terminated
Initiated
2014-07-22
Posted
2014-08-26
Terminated
2015-09-15
Reason
Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
Root cause
Package design/selection
Action
Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.
Quantity
972,125 units
Distribution
Worldwide Distribution.
Lot, serial or model codes
Item Numbers: 00118109006, 00118109008, 00118109010, 00118109506, 00118109508, 00118109510, 00118113008, 00118113010, 00118113012, 00118113014, 00118113508, 00118113510, 00118113512, 00118113514, 00118114008, 00118114010, 00118114012, 00118114014, 00118114508, 00118114510, 00118114512, 00118114514, 00118115006, 00118115008, 00118115010, 00118115012, 00118115014, 00119309006, 00119309008, 00119309010, 00119309012, 00119309014, 00119309506, 00119309508, 00119309510, 00119309512, 00119309514, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514, 00119313008, 00119313010, 00119313012, 00119313014, 00119313508, 00119313510, 00119313512, 00119313514, 00119314008, 00119314010, 00119314012, 00119314014, 00119314508, 00119314510, 00119314512, 00119314514, 00119315008, 00119315010, 00119315012, 00119315014, 00119413008, 00119413508, 00119414008, 00119414508, 00119415008, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514      EXPIRATION DATE PRIOR TO 2024-07
510(k) numbers
["K830196"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register