FDA Medical Device Recalls (openFDA)
Cepheid: Xpert MRSA Catalog GXMRSA-120, Rev. AJ Microbiology: The Cepheid Xpert MRSA assay performed in the GeneXpert Dx System (Xpert MRSA) is a qualitative in vitro diagnostic test designed for rapid detection of Methicillin-Resistant Staphylococcus Aureus (MRSA) from nasal swabs in patients at risk for nasal colonization
- Recall number
- Z-2548-2016
- Recalling firm
- Cepheid
- Product
- Xpert MRSA Catalog GXMRSA-120, Rev. AJ Microbiology: The Cepheid Xpert MRSA assay performed in the GeneXpert Dx System (Xpert MRSA) is a qualitative in vitro diagnostic test designed for rapid detection of Methicillin-Resistant Staphylococcus Aureus (MRSA) from nasal swabs in patients at risk for nasal colonization
- Status
- Terminated
- Initiated
- 2016-07-18
- Posted
- not on file
- Terminated
- 2016-09-30
- Reason
- Internal investigation found the certain lots of MRSA kits have the potential for performance issues at lower positive sample concentrations.
- Root cause
- Nonconforming Material/Component
- Action
- Recall letters dated July 18, 2016 were sent to affected customers on July 19, 2016 by Federal Express. Letters advised users of the affected product and asked that any remaining inventory be destroyed.
- Quantity
- 68 kit for lot 1000037539 and 127 kit s for lot 1000029307.
- Distribution
- US only: PA, DE, FL, SC, ME, CA, IL, NC, TX, WA, NY, TN, FL, OR, OK, MO, MS, GA, NM, IN, OH, MD, VA, CO, AZ, and NC.
- Lot, serial or model codes
- Catalog number: GXMRSA-120, Rev. AJ; Lot numbers: 1000037539 (Cartridge lot 20417) and 1000029307 (Cartridge lot 20419).
- 510(k) numbers
- ["K070462"]
- PMA numbers
- not on file
- Product code
- NQX
- Firm FEI
- 3004530258
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28