Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA

Recall number
Z-2548-2014
Recalling firm
Zimmer, Inc.
Product
Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA
Status
Terminated
Initiated
2014-07-22
Posted
2014-08-26
Terminated
2015-09-15
Reason
Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
Root cause
Package design/selection
Action
Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.
Quantity
972,125 units
Distribution
Worldwide Distribution.
Lot, serial or model codes
Item Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500   Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644
510(k) numbers
["K960278"]
PMA numbers
not on file
Product code
JDI
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register