Zimmer, Inc.: Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO FEMORAL HEAD 45MM DIA ENDO FEMORAL HEAD 46MM DIA ENDO FEMORAL HEAD 47MM DIA ENDO FEMORAL HEAD 48MM DIA ENDO FEMORAL HEAD 49MM DIA ENDO FEMORAL HEAD 50MM DIA ENDO FEMORAL HEAD 51MM DIA ENDO FEMORAL HEAD 52MM DIA ENDO FEMORAL HEAD 53MM DIA ENDO FEMORAL HEAD 54MM DIA ENDO FEMORAL HEAD 55MM DIA ENDO FEMORAL HEAD 57MM DIA ENDO FEMORAL HEAD 60MM DIA ENDO FEMORAL HEAD 63MM DIA ENDO FEM HEAD 41MM ENDO FEM HEAD 42MM ENDO FEM HEAD 43MM ENDO PROS HEAD 44MM ENDO FEM HEAD 45MM ENDO FEM HEAD 46MM ENDO FEM HEAD 47MM ENDO FEM HEAD 48MM ENDO FEM HEAD 49MM ENDO FEM HEAD 50MM ENDO FEM HEAD 51MM ENDO FEM HEAD 52MM ENDO FEM HEAD 53MM ENDO FEM HEAD 54MM ENDO FEM HEAD 55MM ENDO FEM HEAD 57MM ENDO FEM HEAD 60MM ENDO FEM HEAD 63MM UNIPOLAR COMPONENT, 41MM UNIPOLAR COMPONENT, 42MM UNIPOLAR COMPONENT, 43MM UNIPOLAR COMPONENT, 44MM UNIPOLAR COMPONENT, 45MM UNIPOLAR COMPONENT, 46MM UNIPOLAR COMPONENT, 47MM UNIPOLAR COMPONENT, 48MM UNIPOLAR COMPONENT, 49MM UNIPOLAR COMPONENT, 50MM UNIPOLAR COMPONENT, 51MM UNIPOLAR COMPONENT, 52MM UNIPOLAR COMPONENT, 53MM UNIPOLAR COMPONENT, 54MM UNIPOLAR COMPONENT, 55MM UNIPOLAR COMPONENT, 56MM UNIPOLAR COMPONENT, 57MM UNIPOLAR COMPONENT, 58MM UNIPOLAR COMPONENT, 59MM UNIPOLAR COMPONENT, 60MM
- Recall number
- Z-2547-2014
- Recalling firm
- Zimmer, Inc.
- Product
- Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO FEMORAL HEAD 45MM DIA ENDO FEMORAL HEAD 46MM DIA ENDO FEMORAL HEAD 47MM DIA ENDO FEMORAL HEAD 48MM DIA ENDO FEMORAL HEAD 49MM DIA ENDO FEMORAL HEAD 50MM DIA ENDO FEMORAL HEAD 51MM DIA ENDO FEMORAL HEAD 52MM DIA ENDO FEMORAL HEAD 53MM DIA ENDO FEMORAL HEAD 54MM DIA ENDO FEMORAL HEAD 55MM DIA ENDO FEMORAL HEAD 57MM DIA ENDO FEMORAL HEAD 60MM DIA ENDO FEMORAL HEAD 63MM DIA ENDO FEM HEAD 41MM ENDO FEM HEAD 42MM ENDO FEM HEAD 43MM ENDO PROS HEAD 44MM ENDO FEM HEAD 45MM ENDO FEM HEAD 46MM ENDO FEM HEAD 47MM ENDO FEM HEAD 48MM ENDO FEM HEAD 49MM ENDO FEM HEAD 50MM ENDO FEM HEAD 51MM ENDO FEM HEAD 52MM ENDO FEM HEAD 53MM ENDO FEM HEAD 54MM ENDO FEM HEAD 55MM ENDO FEM HEAD 57MM ENDO FEM HEAD 60MM ENDO FEM HEAD 63MM UNIPOLAR COMPONENT, 41MM UNIPOLAR COMPONENT, 42MM UNIPOLAR COMPONENT, 43MM UNIPOLAR COMPONENT, 44MM UNIPOLAR COMPONENT, 45MM UNIPOLAR COMPONENT, 46MM UNIPOLAR COMPONENT, 47MM UNIPOLAR COMPONENT, 48MM UNIPOLAR COMPONENT, 49MM UNIPOLAR COMPONENT, 50MM UNIPOLAR COMPONENT, 51MM UNIPOLAR COMPONENT, 52MM UNIPOLAR COMPONENT, 53MM UNIPOLAR COMPONENT, 54MM UNIPOLAR COMPONENT, 55MM UNIPOLAR COMPONENT, 56MM UNIPOLAR COMPONENT, 57MM UNIPOLAR COMPONENT, 58MM UNIPOLAR COMPONENT, 59MM UNIPOLAR COMPONENT, 60MM
- Status
- Terminated
- Initiated
- 2014-07-22
- Posted
- 2014-08-26
- Terminated
- 2015-09-15
- Reason
- Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
- Root cause
- Package design/selection
- Action
- Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.
- Quantity
- 972,125 units
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- Item Numbers 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500, 00781805700, 00781806000, 00781806300, 00902604101, 00902604201, 00902604301, 00902604401, 00902604501, 00902604601, 00902604701, 00902604801, 00902604901, 00902605001, 00902605101, 00902605201, 00902605301, 00902605401, 00902605501, 00902605701, 00902606001, 00902606301, 451000041, 451000042, 451000043, 451000044, 451000045, 451000046, 451000047, 451000048, 451000049, 451000050, 451000051, 451000052, 451000053, 451000054, 451000055, 451000056, 451000057, 451000058, 451000059, 451000060 EXPIRATION DATE PRIOR TO 2024-01
- 510(k) numbers
- ["K955473"]
- PMA numbers
- not on file
- Product code
- JDI
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28