Beckman Coulter, Inc.: Discrete photometric chemistry analyzer for clinical use.

Recall number
Z-2545-2024
Recalling firm
Beckman Coulter, Inc.
Product
Discrete photometric chemistry analyzer for clinical use.
Status
Open, Classified
Initiated
2024-06-14
Posted
2024-08-09
Terminated
not on file
Reason
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
Root cause
Under Investigation by firm
Action
The recall notification was sent to consignees during the week of 06/14/2024 via email and postal mail. Consignees were instructed to follow the troubleshooting instructions in section 6.4 of the Access 2 Reference Manual and contact the Beckman Coulter representative if the laboratory has experienced repeated reagent carousel motion or QSMC errors. Consignees are asked to respond within 10 days indicating they have received the important communication either via by email or by returning the response form.
Quantity
59 units
Distribution
Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.
Lot, serial or model codes
Catalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJE
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register