FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number/Product Code 6801890, Unique Device Identifier No. 10758750001644, IVD --- Ortho Clinical Diagnostics Inc. The VITROS 5,1 FS Chemistry System with enGEN Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
- Recall number
- Z-2544-2015
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number/Product Code 6801890, Unique Device Identifier No. 10758750001644, IVD --- Ortho Clinical Diagnostics Inc. The VITROS 5,1 FS Chemistry System with enGEN Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
- Status
- Terminated
- Initiated
- 2015-06-30
- Posted
- 2015-09-02
- Terminated
- 2018-05-29
- Reason
- Software anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software Versions 2.2.1 through 2.8. If the user is utilizing User- Configured (Modified) parameters, ALL default settings are restored in the VITROS 5,1 FS Chemistry System from the ADD instead of retaining the user modified parameters.
- Root cause
- Software design
- Action
- The firm, Ortho Clinical Diagnostics, sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letter dated 6/30/15 and response forms to customers via FedEx overnight courier, e-mail (ORTHO PLUS e-Communications) and/or via US Postal Service overnight letter on 6/30/2015. Foreign affiliates were informed by email on 6/30/2015 of the issue and instructed to notify their consignees of the issue and required actions. The letter described the product, problem, and actions to be taken. The customers were instructed to follow the required actions and complete and return the Confirmation of Receipt-Response Required form, no later than 7/8/15, by fax at 1.888.557.3759 or 1.585.453.4110, or scan PDF and email to: ConfirmationAdmin@its.jnj.com. Customers with any additional questions, can contact Customer Technical Services at 1-800-421-3311 at any time.
- Quantity
- Catalog 6801375: Domestic 919; Foreign 1251; Catalog 6801890: Domestic 63, Foreign 147
- Distribution
- Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
- Lot, serial or model codes
- Software Versions 2.2.1 through 2.8; Catalog # 6801375 Serial Numbers: 34000119 through 34002323; Catalog # 6801890 Serial Numbers: J34000133, J34000193, J34000197, J34000212, J34000285, J34000316, J34000318, J34000337, J34000360, J34000420, J34000451, J34000478, J34000495, J34000497, J34000529, J34000538, J34000552, J34000557, J34000590, J34000596, J34000601, J34000631, J34000642, J34000761, J34000799, J34000805, J34000833, J34000851, J34000893, J34000946, J34000947, J34001006, J34001172, J34001218, J34001240, J34001243, J34001248, J34001251, J34001274, J34001297, J34001311, J34001317, J34001345, J34001378, J34001379, J34001389, J34001390, J34001396, J34001422, J34001472, J34001504, J34001569, J34001572, J34001612, J34001626, J34001633, J34001668, J34001755, J34001809, J34001840, J34001850, J34001875, J34001897.
- 510(k) numbers
- ["K031924"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28