FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: Zimmer PERSONA Distal Valgus Alignment Guide, Nonsterile. orthopedic surgical instrument. The device sets the proper Varus/Valgus alignment of a patient's knee prior to the initial femoral resection cut being made.
- Recall number
- Z-2543-2014
- Recalling firm
- Zimmer, Inc.
- Product
- Zimmer PERSONA Distal Valgus Alignment Guide, Nonsterile. orthopedic surgical instrument. The device sets the proper Varus/Valgus alignment of a patient's knee prior to the initial femoral resection cut being made.
- Status
- Terminated
- Initiated
- 2014-07-30
- Posted
- 2014-08-22
- Terminated
- 2015-03-06
- Reason
- Persona Distal Valgus Alignment guide collet locks are seizing in the 'unlocked' position.
- Root cause
- Device Design
- Action
- An Urgent Medical Device Correction letter dated, July 31, 2014, was issued to consignees. The letter identified the affected product, the problem, the risks, and responsibilities. Risk managers and surgeons are to review the notification and the updated surgical technique provided in Attachment 1; as well as, ensure that all affected personnel are aware of the contents. Questions or concerns should be directed to Zimmer customer call center at 1-877-946-2761.
- Quantity
- 2,248 units
- Distribution
- Worldwide Distribution -- United States, including Puerto Rico; and, to the countries of Australia, Austria, Japan, Belgium, Italy, Switzerland, India, Germany, Spain, France, Israel, Netherlands, Saudi Arabia, South Africa & Canada.
- Lot, serial or model codes
- Item Number 42509900400, Lot Number: 62313324, 62415340, 62561031, 62668655, 62313326, 62430425, 62580270, 62695918, 62341561, 62442314, 62590506, 62702686, 62358628, 62473246, 62594059, 62715932, 62368700, 62507846, 62611157, 62728374, 62376588, 62522771, 62624943, 62742362, 62396348, 62530133, 62637095, 62749057, 62401156, 62534170, 62641120, 62757397, 62407514, 62538842, 62662172, 62766074, 62410673, 62542958, & 62668654.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28