Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Zimmer PERSONA Distal Valgus Alignment Guide, Nonsterile. orthopedic surgical instrument. The device sets the proper Varus/Valgus alignment of a patient's knee prior to the initial femoral resection cut being made.

Recall number
Z-2543-2014
Recalling firm
Zimmer, Inc.
Product
Zimmer PERSONA Distal Valgus Alignment Guide, Nonsterile. orthopedic surgical instrument. The device sets the proper Varus/Valgus alignment of a patient's knee prior to the initial femoral resection cut being made.
Status
Terminated
Initiated
2014-07-30
Posted
2014-08-22
Terminated
2015-03-06
Reason
Persona Distal Valgus Alignment guide collet locks are seizing in the 'unlocked' position.
Root cause
Device Design
Action
An Urgent Medical Device Correction letter dated, July 31, 2014, was issued to consignees. The letter identified the affected product, the problem, the risks, and responsibilities. Risk managers and surgeons are to review the notification and the updated surgical technique provided in Attachment 1; as well as, ensure that all affected personnel are aware of the contents. Questions or concerns should be directed to Zimmer customer call center at 1-877-946-2761.
Quantity
2,248 units
Distribution
Worldwide Distribution -- United States, including Puerto Rico; and, to the countries of Australia, Austria, Japan, Belgium, Italy, Switzerland, India, Germany, Spain, France, Israel, Netherlands, Saudi Arabia, South Africa & Canada.
Lot, serial or model codes
Item Number 42509900400, Lot Number: 62313324, 62415340, 62561031, 62668655, 62313326, 62430425, 62580270, 62695918, 62341561, 62442314, 62590506, 62702686, 62358628, 62473246, 62594059, 62715932, 62368700, 62507846, 62611157, 62728374, 62376588, 62522771, 62624943, 62742362, 62396348, 62530133, 62637095, 62749057, 62401156, 62534170, 62641120, 62757397, 62407514, 62538842, 62662172, 62766074, 62410673, 62542958, & 62668654.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register