FDA Medical Device Recalls (openFDA)
Wright Medical Technology, Inc.: CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x 12 mm, REF CCP-L3012; 3.0 x 14 mm, REF CCP-L3014; 3.0 x 16 mm, REF CCP-L3016; 3.0 x 18 mm, REF CCP-L3018; 3.0 x 20 mm, REF CCP-L3020; 3.0 x 22 mm, REF CCP-L3022; 3.0 x 24 mm, REF CCP-L3024; 3.0 x 26 mm, REF CCP-L3026; 3.0 x 28 mm, REF CCP-L3028; 3.0 x 30 mm, REF CCP-L3030; 3.0 x 32 mm, REF CCP-L3032; 3.0 x 34mm, REF CCP-L3034; 3.0 x 36 mm, REF CCP-L3036; 3.0 x 38 mm, REF CCP-L3038; 3.0 x 40 mm, REF CCP-L3040; 3.5 x 8 mm, REF CCP-L3508; 3.5 x 10 mm, REF CCP-L3510; 3.5 x 12 mm, REF CCP-L3512; 3.5 x 14mm, REF CCP-L3514; 3.5 x 16 mm, REF CCP-L3516; 3.5 x 18 mm, REF CCP-L3518; 3.5 x 20 mm, REF CCP-L3520; 3.5 x 22 mm, REF CCP-L3522; 3.5 x 24 mm, REF CCP-L3524; 3.5 x 26 mm, REF CCP-L3526; 3.5 x 28 mm, REF CCP-L3528; 3.5 x 30 mm, REF CCP-L3530; 3.5 x 35 mm, REF CCP-L3532; 3.5 x 34 mm, REF CCP-L3534; 3.5 x 36 mm, REF CCP-L3536; 3.5 x 38 mm, REF CCP-L3538; 3.5 x 40 mm, REF CCP-L3540 The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.
- Recall number
- Z-2541-2015
- Recalling firm
- Wright Medical Technology, Inc.
- Product
- CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x 12 mm, REF CCP-L3012; 3.0 x 14 mm, REF CCP-L3014; 3.0 x 16 mm, REF CCP-L3016; 3.0 x 18 mm, REF CCP-L3018; 3.0 x 20 mm, REF CCP-L3020; 3.0 x 22 mm, REF CCP-L3022; 3.0 x 24 mm, REF CCP-L3024; 3.0 x 26 mm, REF CCP-L3026; 3.0 x 28 mm, REF CCP-L3028; 3.0 x 30 mm, REF CCP-L3030; 3.0 x 32 mm, REF CCP-L3032; 3.0 x 34mm, REF CCP-L3034; 3.0 x 36 mm, REF CCP-L3036; 3.0 x 38 mm, REF CCP-L3038; 3.0 x 40 mm, REF CCP-L3040; 3.5 x 8 mm, REF CCP-L3508; 3.5 x 10 mm, REF CCP-L3510; 3.5 x 12 mm, REF CCP-L3512; 3.5 x 14mm, REF CCP-L3514; 3.5 x 16 mm, REF CCP-L3516; 3.5 x 18 mm, REF CCP-L3518; 3.5 x 20 mm, REF CCP-L3520; 3.5 x 22 mm, REF CCP-L3522; 3.5 x 24 mm, REF CCP-L3524; 3.5 x 26 mm, REF CCP-L3526; 3.5 x 28 mm, REF CCP-L3528; 3.5 x 30 mm, REF CCP-L3530; 3.5 x 35 mm, REF CCP-L3532; 3.5 x 34 mm, REF CCP-L3534; 3.5 x 36 mm, REF CCP-L3536; 3.5 x 38 mm, REF CCP-L3538; 3.5 x 40 mm, REF CCP-L3540 The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.
- Status
- Terminated
- Initiated
- 2015-07-21
- Posted
- 2015-09-01
- Terminated
- 2018-08-01
- Reason
- These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.
- Root cause
- Employee error
- Action
- Wright sent an Urgent Medical Device Recall letter dated July 20, 2015, to all affected customers. Customers were asked to complete the attached response form confirming receipt of the notice and the status of any products on hand, and return a copy to Wright by email (complaints@wmt.com or fax (901)867-7401. Also, if customers have any product in their inventory the product should be sent back to Arlington, TN (ALM). For questions regarding this recall call 901-290-5662.
- Quantity
- 107 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- CrossCHECK Locking Screw with the following codes: REF CCP-L3008: Lot Number 35000; REF CCP-L3010: Lot Number 35000; REF CCP-L3012: Lot Number 35000-6, 35000; REF CCP-L3014: Lot Number 35000-8, 35000; REF CCP-L3016: Lot Number 35000; REF CCP-L3018: Lot Number 35000; REF CCP-L3020: Lot Number 35000; REF CCP-L3022: Lot Numbers: 35000-16-1, 35000; REF CCP-L3024: Lot Number 35000; REF CCP-L3026: Lot Number 35000; REF CCP-L3028: Lot Number 35000; REF CCP-L3030: Lot Number 35000; REF CCP-L3032: Lot Number 35000; REF CCP-L3034: Lot Number 35000; REF CCP-L3036: Lot Number 35000; REF CCP-L3038: Lot Number 35000; REF CCP-L3040: Lot Number 35000; REF CCP-L3508: Lot Number 35004; REF CCP-L3510: Lot Numbers: 35004-3, 35004; REF CCP-L3512: Lot Numbers: 35004-6, 35004; REF CCP-L3514: Lot Numbers: 35004-7, 35004; REF CCP-L3516: Lot Number 35004; REF CCP-L3518: Lot Numbers: 35004-12, 35004-12-1, 35004; REF CCP-L3520: Lot Numbers: 35004-13, 35004; REF CCP-L3522: Lot Numbers: 35004-16-1, 35004; REF CCP-L3524: Lot Numbers: 35004-18-1, 35004; REF CCP-L3526: Lot Number 35004; REF CCP-L3528: Lot Number 35004; REF CCP-L3530: Lot Number 35004; REF CCP-L3532: Lot Number 35004; REF CCP-L3534: Lot Number 35004; REF CCP-L3536: Lot Number 35004; REF CCP-L3538: Lot Number 35004; REF CCP-L3540: Lot Number 35004
- 510(k) numbers
- ["K132594"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 3010667733
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28