Skip to the content

FDA Medical Device Recalls (openFDA)

Limacorporate S.p.A: SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.

Recall number
Z-2540-2015
Recalling firm
Limacorporate S.p.A
Product
SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.
Status
Terminated
Initiated
2015-08-14
Posted
2015-09-02
Terminated
2016-02-26
Reason
The dimensions of the mating features of the recalled products are out of specification causing the device to be difficult to assemble.
Root cause
Nonconforming Material/Component
Action
Lima Corporate sent an Urgent Attention Recall letter on August 11, 2015, to US distributors and Sales Agents. The affected customers were instructed to examine their inventory, return the affected lots to the recalling firm, and complete a response form.
Quantity
93 units
Distribution
US Distribution to the states of: OH, TX, NC, NY, CA, FL, KS, AL, MI, MT, CO, MO, NV, WA, MN and LA., and Internationally to Australia.
Lot, serial or model codes
Lot Numbers: 14AA146 and 14AA153
510(k) numbers
["K100858"]
PMA numbers
not on file
Product code
HSD
Firm FEI
3008021110
Firm city
San Daniele Del Friuli
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register