Skip to the content

FDA Medical Device Recalls (openFDA)

Bound Tree Medical: Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu King LTS-D" Supraglottic Airway.

Recall number
Z-2532-2020
Recalling firm
Bound Tree Medical
Product
Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu King LTS-D" Supraglottic Airway.
Status
Terminated
Initiated
2020-05-14
Posted
not on file
Terminated
2023-12-07
Reason
Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit
Root cause
Process control
Action
The firm, Bound Tree, issued "URGENT NOTICE OF RECALL" letter by mail on May 14, 2020 and Bound Tree Customer Service and followed with email or telephone calls. Letter states reason for recall and action to take: Please immediately physically review your inventory of this kit and lot number for the affected syringe . Should you determine your inventory contains affected product, please quarantine affected kits immediately and contact Boundtree customer care to return product for replacement or credit. Questions pertaining to the recall, you may contact Boundtree customer care at 800-533-0523.
Quantity
167 kits
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Number: ASM0025699
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAE
Firm FEI
3010792353
Firm city
Dublin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register