Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444

Recall number
Z-2530-2026
Recalling firm
Zimmer, Inc.
Product
Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444
Status
Open, Classified
Initiated
2026-05-20
Posted
2026-06-23
Terminated
not on file
Reason
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE RECALL CORRECTION notification letter dated 5/26/26 was sent to customers. Distributor Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. 3. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons who are current users of the affected products that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Risk Manager Responsibilities: 1. Review this correction notice and ensure that affected personnel are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 3. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, y
Quantity
11 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot Code: 110040444 ; UDI-DI (01)00889024681903(17)350531(10)097487 ; Lot Number 097487
510(k) numbers
["K241043"]
PMA numbers
not on file
Product code
PHX
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register