Siemens Medical Solutions USA, Inc: Artis Pheno. Image-Intensified Flouroscopic X-Ray System.

Recall number
Z-2529-2025
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis Pheno. Image-Intensified Flouroscopic X-Ray System.
Status
Open, Classified
Initiated
2025-08-12
Posted
2025-09-05
Terminated
not on file
Reason
Limited system movements after startup .
Root cause
Software design
Action
Consignees were mailed an URGENT: MEDICAL DEVICE CORRECTION notification dated 8/12/25. The notification informs consignees that the recall issue will be corrected via release of an Update Instruction AX032/25/S by appointment, which can be arranged by contacting 1-800-888-7436. Consignees are to disseminate the recall notification to all users of affected devices within their organization and to where product may be further distributed. Consignees are to return the completed recall response form via email to recallsandrefusals.team@siemens-healthineers.com or by certified mail.
Quantity
20 units
Distribution
US Nationwide distribution in the states of AL, AZ, FL, KY, MA, MI, NY, OR, A, TN, TX & UT.
Lot, serial or model codes
Model No. 10849000; UDI: 04056869046877; Serial No. 164184, 164183, 164071, 164072, 164027, 164026, 164094, 164019, 164161, 164152, 164069, 164086, 164084, 164158, 164188, 164105, 164080, 164011, 164100, 164097.
510(k) numbers
["K241572"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register