Siemens Medical Solutions USA, Inc: Artis Pheno. Image-Intensified Flouroscopic X-Ray System.
- Recall number
- Z-2529-2025
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Artis Pheno. Image-Intensified Flouroscopic X-Ray System.
- Status
- Open, Classified
- Initiated
- 2025-08-12
- Posted
- 2025-09-05
- Terminated
- not on file
- Reason
- Limited system movements after startup .
- Root cause
- Software design
- Action
- Consignees were mailed an URGENT: MEDICAL DEVICE CORRECTION notification dated 8/12/25. The notification informs consignees that the recall issue will be corrected via release of an Update Instruction AX032/25/S by appointment, which can be arranged by contacting 1-800-888-7436. Consignees are to disseminate the recall notification to all users of affected devices within their organization and to where product may be further distributed. Consignees are to return the completed recall response form via email to recallsandrefusals.team@siemens-healthineers.com or by certified mail.
- Quantity
- 20 units
- Distribution
- US Nationwide distribution in the states of AL, AZ, FL, KY, MA, MI, NY, OR, A, TN, TX & UT.
- Lot, serial or model codes
- Model No. 10849000; UDI: 04056869046877; Serial No. 164184, 164183, 164071, 164072, 164027, 164026, 164094, 164019, 164161, 164152, 164069, 164086, 164084, 164158, 164188, 164105, 164080, 164011, 164100, 164097.
- 510(k) numbers
- ["K241572"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28