FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: M3150 Information Center Local Database; Models: 862249, 865001, 865003, 865004, 865007, 865009, 865011, 865012, 865091, 865092, 865093, 865094, 865095, 865096, 865098, 865099, 865102, 865104, 865105, 865125, 865127. The intended use of the Information Center Software is to display physiologic waves, parameters and trends, formal data for strip recordings and printed reports and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
- Recall number
- Z-2528-2017
- Recalling firm
- Philips Electronics North America Corporation
- Product
- M3150 Information Center Local Database; Models: 862249, 865001, 865003, 865004, 865007, 865009, 865011, 865012, 865091, 865092, 865093, 865094, 865095, 865096, 865098, 865099, 865102, 865104, 865105, 865125, 865127. The intended use of the Information Center Software is to display physiologic waves, parameters and trends, formal data for strip recordings and printed reports and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
- Status
- Terminated
- Initiated
- 2010-12-17
- Posted
- 2017-06-13
- Terminated
- 2017-08-01
- Reason
- Potential for early component failure.
- Root cause
- Component design/selection
- Action
- Philips sent an Urgent Medical Device Correction letter dated December 17, 2010, to all affected consignees. Philips will replace defective system boards in Hewlett Packard model rp5700 PCs sold with IntelliVue Information Centers to correct the problem. Philips Healthcare representatives will inspect all affected units, and any unit identified to have a defective system board will have its system board replaced at no charge to the customer. Philips is asking the following of customers in the Action to be Taken by Customer/User section of the FSN: Observe your system for any of the symptoms described in the Problem Description section above. If symptoms are noted, please contact your local Philips Healthcare Customer Service representative or the Customer Care Solutions Center at (800) 722-9377 # 3, # 1. Do not rely exclusively on the audible alarm system for patient monitoring. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment. For questions regarding this recall call 978-659-3000.
- Quantity
- 607 units
- Distribution
- Worldwide Distribution - US (Nationwide) OUS: CANADA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL CHILE CHINA COLOMBIA CZECH REPUBLIC FINLAND FRANCE GERMANY GREECE INDIA INDONESIA IRELAND ITALY JAPAN KENYA KOREA, REPUBLIC OF NETHERLANDS PHILIPPINES POLAND SPAIN SWEDEN SWITZERLAND THAILAND TURKEY UNITED KINGDOM
- Lot, serial or model codes
- This action affects certain customers who were shipped an affected Philips Intellivue Information Center unit between June 9, 2008 and May 31, 2009. Affected products have a PC serial number range between 2UA8300XYZ and 2UA84510QW. However, only a subset of these products are affected.
- 510(k) numbers
- ["K062271"]
- PMA numbers
- not on file
- Product code
- MHX
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28