Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical Inc.: Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow stop, yellow-striped tubing, clamp and female NRFit connector. Yellow stopper included 100mL", Product Code 21-7600-24

Recall number
Z-2527-2025
Recalling firm
ICU Medical Inc.
Product
Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow stop, yellow-striped tubing, clamp and female NRFit connector. Yellow stopper included 100mL", Product Code 21-7600-24
Status
Open, Classified
Initiated
2025-08-06
Posted
2025-09-04
Terminated
not on file
Reason
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Root cause
Under Investigation by firm
Action
ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 08/06/2025 via registered mail and email. The notice explained the issue, potential risk can requested the following: Customer Required Actions: Please complete the following actions below. 1. Review your inventory of CADD Medication Cassette Reservoirs to confirm if any of the affected lots are within your possession. 2. Quarantine the affected product and destroy or discard it immediately following your institution s process for destruction or discarding. 3. Return the attached Customer Response Form to ICUMedical8176@sedgwick.com within ten (10) days of receipt to acknowledge your understanding of this notification. 4. DISTRIBUTORS: Immediately forward this notice to all users to whom you distributed these products. Request that they complete the Customer Response Form and return it to ICUMedical8176@sedgwick.com. For further inquires: - To report adverse events or product complaints: Global Complaint Management: productcomplaints@icumed.com; 1-(866)-216-8806 - Additional information or to request replacement product: Customer Service: customerservice@icumed.com; 1-(800)-258-5361 - Questions about this communication: Field Action Processing: ICUMedical8176@sedgwick.com; 1-(844)-861-6220
Quantity
15546 units
Distribution
Worldwide distribution.
Lot, serial or model codes
UDI/DI 10610586044007, Lot Numbers: 6037784, 6054015
510(k) numbers
not on file
PMA numbers
not on file
Product code
FPA
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register